Free Essay

Calibration

In:

Submitted By aziz6193
Words 2383
Pages 10
ABSTRACT
This experiment will examine the concentrations of Zinc in several "samples." We will be using a Zinc lamp on the flame atomic absorption spectrophotometer. The lab assistant will demonstrate the software and the procedures for preparing the instrument.
Each group will run a series of calibration standards to generate a calibration curves. The concentrations are given below. Then, one of the "unknown" samples (A or B) should be run. The printer will generate a report of the concentration of Zinc in the sample. We have to be sure which sample we are used.
Our laboratory report should describe the functions of the AAS and describe the procedures for preparing the instrument for use. The report should include a printout graph of the calibration curve that we have made. We also needed to indicate the data points for the calibration standards and our sample.

INTRODUCTION
Atomic absorption spectroscopy (AAS) determines the presence of metals in liquid samples. Metals include Fe, Cu, Al, Pb, Ca, Zn, Cd and many more. It also measures the concentrations of metals in the samples. Typical concentrations range in the low mg/L range.
In their elemental form, metals will absorb ultraviolet light when they are excited by heat. Each metal has a characteristic wavelength that will be absorbed. The AAS instrument looks for a particular metal by focusing a beam of UV light at a specific wavelength through a flame and into a detector. The sample of interest is aspirated into the flame. If that metal is present in the sample, it will absorb some of the light, thus reducing its intensity. The instrument measures the change in intensity. A computer data system converts the change in intensity into an absorbance.
As concentration goes up, absorbance goes up. The researcher can construct a calibration curve by running standards of various concentrations on the AAS and observing the absorbances. In this lab, the computer data system will draw the curve for you! Then samples can be tested and measured against this curve.
We describe the motion of macroscopic objects using Newton's laws of physics. We can easily measure velocity, acceleration, force, kinetic and potential energies as well as quantify gravitational effects. However, when we look at atomic particles (i.e., protons, electrons, neutrons, etc.), we can no longer describe motion and energy using Newtonian physics. The characteristics of atomic particles are described using the theorems of quantum mechanics.
Quantum mechanics--or quantum chemistry--describes the geometry of atoms and molecules in terms of complex mathematical expressions. It also describes the relative states of atomic matter. The atomic absorption spectrometer uses the principals of quantum chemistry to detect the presence of certain metals (i.e., iron, aluminium, copper, etc.) and determines the concentration of those metals in samples.
All atoms and their components have energy. The energy level at which an atom exists is referred to as its state. Under normal conditions, atoms exist in their most stable states. We refer to that most-stable level as the ground state. Al though we cannot measure the precise energy state for an atom, we can usually measure changes to its energy relative to its ground state.
Certain processes can change the energy state for an atom. For example, adding thermal energy (heat) can cause an atom to increase to a higher energy state. This change in energy is written as DE. We refer to energy states which are higher than the ground state as excited states. In theory, there are infinite excited states; however there are decreasing numbers of atoms from a population that reach higher excited states.
The laws of quantum mechanics tell us that atoms do not increase their energy levels gradually. An atom goes directly from one state to another without going through intermediates. We refer to these "quantum leaps" as transitions. The transition from the ground state (written as Eo) to the first excited state (E1) requires some form of energy input. This energy is absorbed by the atom. That energy absorption is equal to DE0® 1. When this energy absorption takes place in the presence of ultraviolet light, some of that light will be absorbed. This uv absorption occurs at a specific wavelength.
Each element in the periodic table will have a specific D E that will absorb a specific wavelength of uv light. The relationship between the energy transition and the wavelength (l) can be described by:
D E=h/ l
Where h is Planck's constant. Atomic absorption uses this relationship to determine the presence of a specific element based on absorption in a specific wavelength. For example, calcium absorbs light with a wavelength of 422.7 nm. Iron absorbs light at 248.3 nm.

APPARATUS
Volumetric flasks, Pipette and Atomic Absorption Spectrometer
CHEMICALS
Zinc Sulphate, Deionised water and Water sample
PROCEDURE
A. PREPARATION OF STOCK SOLUTION (1000 PPM Zn) 100 ml of 1000 ppm Zn solution was prepared by weighing 0.4407 g ZnSO4.7H2O. Then the salt was dissolved in some deionised water. Next the solution was quantitatively transferred into a 100 ml volumetric flask. Deionised water was added to the mark. Lastly the exact concentration (ppm) of Zn was calculated in the stock solution.
B. DETERMINATION OF Zn USING THE LINEAR CALIBRATION METHOD Standards was prepared by calculating the volume of the stock solution required to prepared a series of 50 ml standards of Zn (0.1, 0.2, 0.3, 0.4, 0.5 ppm). A serial dilution was performed by pipetting the calculated volumes into 50 ml flasks. Deionised water was added to the mark. Next 10 ml of the unknown sample was pipeted into a 50 ml volumetric flask. Distilled water was added to the mark. After that, absorbance was measured by recording the absorption readings of each samples followed by the sample. Distilled water was used as Blank. Lastly Zn in the sample was determined by using ‘Excel Spreadsheet’. Calibration graph of absorbance versus concentration of Zn was plotted by using ‘Excel Spreadsheet’. The linear equation and the R² value was generated. The concentration of Zn in the sample was determined by using the linear equation of the calibration graph.
C. DETERMINATION OF Zn USING THE STANDARD ADDITIONS METHOD Standards and sample were prepared by calculating the volume of the Zn stock solution needed to prepare a series of five 100 ml standards with concentrations of (x+0.00), (x+0.25), (x+0.50), (x+0.75), (x+1.00) ppm. The appropriate volume of stock solution as required was pipetted into each 100 ml volumetric flask. Next 20 ml of the water sample was pipetted into each flask. Deionised water was added to the mark. After that, absorbance was measured. Distilled water was used as Blank. Calibration graph of absorbance against concentration of Zn was plotted by using ‘Excel Spreadsheet’. The concentration of Zn (x) from negative intercept of the x-axis was determined.

Results and Calculation:
Preparation of Stock Solution (10 ppm Zn)
Weight of Zinc Sulphate for standard: 0.4407
Dilution Factor: 1000/10: 100
Concentration of Zinc: 10ppm Determination of Zn using the Linear Calibration Curve Method
Since Blank equal to zero, Absorbance Reading already calibrated Sample | Concentration(ppm) | Calibratted Absorbance Reading | Blank | 0.00 | 0.0000 | Standard 1 | 0.10 | 0.0413 | Standard 2 | 0.20 | 0.0851 | Standard 3 | 0.30 | 0.1556 | Standard 4 | 0.40 | 0.1697 | Standard 5 | 0.50 | 0.2541 |
Table 1: Calibrated reading of the blank and standard sample of zinc concentration
Atomic Absorbance for sample A: -0.0610
Linear Method
Using the linear equation,
: Atomic Absorbance Value
: Concentration of Zinc Sample (diluted)

Rearrange the equation:
: -0.0610

= -0.06276
Sample A: 10 ml in 50 ml volumetric flask
Dilution Factor: 50/10 ml = 5
So, the concentration of zinc in the sample

=0.00 ppm
Determination of Zn using the Standard Addition method
Since Blank equal to zero, Absorbance Reading already calibrated Sample | Concentration | Standard Concentration added to sample (ppm) | Blank | 0.0000 | 0.00 | X+0.00 standard | 0.0208 | 0.00 | X+0.25 standard | 0.2655 | 0.25 | X+0.50 standard | 0.4477 | 0.50 | X+0.75 standard | 0.7039 | 0.75 | X+1.00 standard | 1.0186 | 1.00 |
Table 2: Calibrated reading of the blank and standard concentration added to the zinc sample
Calibrated linear standard Method
Sample A: 20 ml in 100 ml volumetric flask
Dilution Factor: 100/20 ml = 5
: Atomic Absorbance Value
: Standard Added + Zinc Sample (diluted)
So, the concentration of Zinc Sample when extrapolated at y=0:

, thus
So, the concentration of zinc in the sample
Standard Addition method

Discussion In this experiment, the concentration of zinc in the sample was being determined. There are two type of method applied here. First method is using linear calibration graph and second method is using standard addition method. So, by comparing the concentration obtained from two methods the concentration can be checked whether correct or not. By using atomic absorption spectroscopy (AAS), the concentration of metal element or mineral in a sample can be determined. This technique use absorption or emission spectrometry to assess the concentration of analytes in sample. To conduct the experiment, a range of standards must be prepared containing varying amounts of the analyte. The amount of analyte present in the sample can be calculated by plotting a calibration curve of signal against amount of analyte in the sample analyzed. The linear equation can be generated and the equation can be used to calculate amount of zinc in the unknown water sample A. The second method was known as standard addition method.; this method often used to reduce the problems of matrix effects. To find the amount of zinc from the water sample, certain amount of sample was mixed with known standard concentration. From the graph, a line can be extrapolate and value of x represent the concentration of the zinc in the sample at y=0. In part A of the experiment, 100 ml of 1000 ppm Zn solution were made by weighing 0.4407 g Zn SO4.7H2O in 100 ml volumetric flask. To make the process of producing the standards, 1000 ppm Zn were diluted to 10 ppm by taking 10 ml of 1000 ppm Zn solution made into another 100 ml volumetric flask. To determine the concentration of zinc in water sample, the standard sample and unknown water sample A must be analyzed under atomic absorption spectrometer. The standard sample series was ranged from 0.1 ppm to 0.5 ppm. The reading of the standard sample was tabulated into table 1.Based on the data obtain a calibration graph can be plotted in figure 1. Since the blank does not show any reading, thus the instrument already calibrated and the point can be used without subtract the blank value. The concentration must be measured from low to high to avoid previous Zinc concentration affect the sample afterwards. Based on the graph of Linear equation of the calibration graph 1, a linear equation was generated. By using this equation, y = 0.9736x + 0.045, x equal to concentration of the analyte in the zinc water sample and y is the reading of the absorbance value. So, from the calculation, the calculated concentration of zinc in water sample equal to -0.06276. This is not the real concentration of the sample, the concentration of the sample must be times with it dilution factor 5 times giving value of -0.3138. However, this value is incorrect since the concentration cannot be negative. So, it can be conclude that the concentration of the sample should be zero ppm or less then the limit of detection of the instrument. The second method is through the standard addition technique. The advantage of using this technique was that the effect of the matrix on the analyte can be reduced. As the results, more accurate result can be obtained. To determine the concentration of zinc in the water sample, 20 ml of water sample A was mix known amount of standard zinc sample. The series ranged from 0 to 1.00 ppm. The result was tabulated into table 2 and a calibrated graph 2 was plotted based on the data in table 2. The analysis of the data of the second method give that the concentration of the zinc in the water sample was -0.0578 after times with its dilution factor of 5. The concentration was obtained by measuring the value of x when extrapolated. Since x cannot be reading from the graph 2 easily, the generated linear equation was used instead by set y = 0. Based on the equation, y = 0.4932x – 0.0057, x can be find by equating y = 0. So, the value of x was 0.01156 but the concentration equal to –x. So, the concentration of the sample after dilution equal to -0.01156 and in real sample equal -0.0578. Same as method 1, the result give negative value, since concentration cannot be negative, so the concentration of zinc in water sample A must be zero ppm or less than the detection limit of the instrument. There are few precaution need to be taken to avoid errors during the experiment. Firstly, the concentration of the standard sample must be tested from lower concentration to high concentration. Meanwhile the sample must be tested first to check whether the sample reading within the standard range. Secondly, the sample must be diluted first if the concentration is higher than the standard range. The concentration of the sample A calculated equal to zero. However, the concentration might not accurate because of the instrument. If the concentration of the zinc was to low pass the limit of detection the concentration cannot be measured.
Conclusion
The concentration of sample A was identified to zero ppm. By using the same method atomic absorption spectrometer give out negative value. For linear graph method it give -0.3138 ppm and standard addition method it gives -0.0578 ppm. The concentration of sample A is not measured correctly either the problem was the sample or the instrument.
References
1- Galen Wood Ewing (1997), Analytical Instrumental handbook 2nd ed. Pg 257 – 266, USA 2- R.S. Khandpur (2006), Handbook of Analytical Instruments, Pg 32 – 34, Tata Mc Graw Hill

Similar Documents

Free Essay

Uv Calibration

...UV-Visible Spectrophotometer Calibration S.ZALI M.Sc. Dept. of Analytical chemistry Razi vaccine serum research institute S.Zali@rvsri.ir 2008 The primary objective of GLP is to ensure the generation of high quality data. Essential to this is the ability to produce estimates for the measurement uncertainty associated with calibrations. V The spectrophotometer has become the workhorse for high throughput routine testing for a wide variety of analysts in chemical and biochemical analysis laboratories. It is important for accuracy and reproducibility in measurement that this instrument is correctly set up and calibrated. Increasing Wavelength 50,000 Gamma X-ray 12,820 UV 200 4000 NIR 380 780 IR 2500 400 cm-1 FIR Micro Radio 25 000 nm Increasing Energy K-shell electrons Outer-shell electrons Molecular vibrations Molecular Rotation Conventional Spectrophotometer: The factors that must be calibrated: Control of wavelengths Control of absorbance Limit of stray light Resolution Spectral slit width Cell Control of wavelengths: Wavelength standards: The ideal wavelength standard would have very narrow and very well defined peaks. Control of wavelengths: •Solution of holmium(III) in perchloric acid: 240– 640 nm. •Holmium oxide filters: 280–640 nm. •Didymium filters: 400–750 nm. •Using of spectral lines in low-pressure discharge lamps: 656.1 nm produced by...

Words: 982 - Pages: 4

Free Essay

Sdfb

...THE POTENTIOMETER: Internal Resistance of a Test Cell OBJECT To calibrate a one meter slide wire potentiometer using a standard cell and then to use this potentiometer to measure the emf of a test cell. The terminal voltage of the same test cell is then measured as different load resistors are connected across the test cell and these data are used to determine the internal resistance of the test cell. THEORY The electromotive force (emf) of a cell is its terminal voltage when no current is flowing through it. The terminal voltage of a cell is the potential difference between its electrodes. A voltmeter cannot be used to measure the emf of a cell because a voltmeter draws some current from the cell. To measure a cell's emf a potentiometer is used since in a potentiometer measurement no current is flowing. It employs a null method of measuring potential difference, so that when a balance is reached and the reading is being taken, no current is drawn from the source to be measured. Figure 1. This is the basic circuit diagram for a potentiometer.  Point C is the sliding contact which can be adjusted for zero current deflection through the galvanometer. In this method (refer to Figure 1) a uniform, bare slide wire AB is connected across the power supply. If you were to connect a voltmeter between the + power supply terminal and point A you would measure essentially zero volts. If you were to now connect the voltmeter between the + power supply and point B you would measure...

Words: 1527 - Pages: 7

Free Essay

Joel Hills Paper

...Wilson Electronics Cellular Booster Database Requirements For Cellular Booster O-14828 Version 1.06 Presented to: Wilson Electronics Approval Name Title Signature Date Confidential Wilson Electronics Cellular Booster Database Requirements - v1.06 O-14828 Change History Version # 1.00 1.01 1.02 1.03 Reason Author Date 2014/04/11 2014/04/16 2014/04/21 2014/04/22 Initial Release Condensed parameters Consensus reached on majority of parameters Added specifics about DB connection, changed AGC Test Parameters Moved some configuration items to test limit sections Al Braga Al Braga Al Braga Al Braga Al Braga 1.04 2014/04/30 1.05 1.06 Merged RF Test Sections, added error, clarified Al Braga some parameter types. Removed all Station Name inputs Al Braga 2014/05/14 2014/05/15 Page 2 of 14 Project File: Cellular Booster Database Requirements.docx ©Averna - Confidential Wilson Electronics Cellular Booster Database Requirements - v1.06 O-14828 Table of content Purpose and Scope ....................................................................................................................................... 4 Considerations .............................................................................................................................................. 4 Database Connectivity .................................................................................................................................. 5 Database...

Words: 2582 - Pages: 11

Free Essay

Quality One Call at a Time

...RUNNING HEAD: Quality One Call at a Time Quality One Call at a Time May 30, 2006 Quality scores in many call centers are on the decline. The main goal of a call center is to provide excellent customer service while delivering accurate information in a timely manner. “Fast is fine, but accuracy is everything.” Wyatt Earp. Wyatt Earp may not have said this quote in reference to managing quality at call centers, but it carries the same relevance. Management at call centers tends to focus on other issues instead of the accuracy of the information that the CSR is relaying to the customer. Thus value is misplaced and the quality at call centers starts to head on a downward path. One issue that could drive quality scores down is ineffective training. At a call center here in Jacksonville, management decided to pull CSRs out of training after the second day to answer calls because the queue was high. This interrupted the training process and added days to the training schedule. The trainer manager was powerless to prevent this from happening. This happened 6 times over the course of the proposed three-week training calendar. Call center trainers must ensure that their trainees’ learning is not hindered by management’s goal to reduce queue times. The constant yo-yo handling of the CSRs resulted in training not being delivered on all topics, which resulted in low quality scores. This has not been a high topic of focus for most call center management staff. ...

Words: 836 - Pages: 4

Free Essay

Zghgfhgf

...Math – Final Exam Vet 110 Name: _______________________ Date: _______________________ Good luck! ( Identify the highest value: 1) 3.7 4.1 2.9 2) 5.4 6.2 4.7 3) 1.2 1.5 1.7 Add the following decimals: 4) 1.35 + 2.42 = ________ 5) 2.8 + 0.03 = ________ Subtract the following decimals: 6) 2.3 – 1.45 = ________ 7) 10.25 – 1.47 = ________ Multiply the following: 8) 1.37 x 0.02 = ________ 9) 0.31 x 0.04 = ________ Eliminate the decimal: 10) 5.25 = ________ 0.3 Solve the following equations; express your answer to the nearest tenth. 11) 12 x 500 x 2000 x 1 = ________ 1 2700 1 60 12) 0.35 x 2.5 = ________ 7. 0.7 13) How many tablets will be needed to give the following dose? The tablets are labeled 1.2 mg. The order is for 0.6 mg. ________ tablet(s) 14) Reduce the following as far as possible: 124 76 = ___________________________ 15) Express the following division to the nearest tenth: 5.7 1.46 = __________________________ 16) Solve the following equation. Express answer to the nearest tenth. 30 x 50 x 10 x 415 1 550 1250 1 = ____________________ Write the following metric measures using their abbreviations and notation rules. 17) Four hundredth of a gram ___________ 18) ...

Words: 648 - Pages: 3

Premium Essay

Surface Well Test

...SURFACE WELL TESTING By Karrym Rebay Vol. 1 Preface This is a paper that has been produced to help you better understand and work with the equipment you are familiar with and presently operating on your particular Well Test Unit. It will give you more knowledge about, not only the operation, but also the maintenance and calibration techniques that will help you become more familiar with your own equipment. Paper Content 1. PAPER DESCRIPTION 2. IDENTIFYING A STANDARD WELLTEST PACKAGE AND ITS COMPONENTS 3. EQUIPMENT OPERATION AND FLOWING CONDITIONS 4. SAFETY AROUND YOUR EQUIPMENT 5. CALIBRATION AND MAINTENANCE 6. CALCULATING FLOW RATES 7. GLOSSARY – Exploration & Production Terms 8. CONVERSIONS & TABLES Paper Description This paper describes the specific conditions under which well tests must be performed, lists the surface testing equipment used to perform these well tests, summarizes how this equipment is used to collect samples at the surface and lists several examples that influence the layout of surface equipment. A reservoir test can only be performed under certain conditions. This means the reservoir must be exposed to a disturbance that will cause the reservoir pressure to change. This pressure change, when recorded and interpreted along with the measured flow rates, will give us information about well and reservoir parameters and geometry. A pressure disturbance...

Words: 873 - Pages: 4

Premium Essay

Examples of Possible Exam Questions Sustainable Business and Ethics Course

...We expect that you would be able to find at least 10 to 15 standards for your product. For each standard, you should be able to indicate the answers to each column of the CSR standards inventarisation excel sheet. Namely: Name of the standards you have found relevant to your chosen product Explanation of the purpose of the standard Applicable countries Applicable trade region(s) Application to the chosen product Website address(es) Possible market consequences for the supply chain Example: Name of the standards you have found relevant to your chosen product: Florverde Explanation of the purpose of the standard: Florverde is an initiative of ASOCOFLORES the Colombian association of flowers exporters. The aim of the combined action is to satisfy the requirements of Colombian’s flower 1st importer market the USA. The Florverde certification set up in 1996 enables the thousands of workers involved in the production of export flowers to live and work in acceptable conditions. The certification secures the main USA market and maintain best sustainable practices through annual audits. The main areas for standards include: basic labor regulations, personnel management, work and social welfare, water and irrigation, intergrated pest and disease management, traceability and records. Applicable countries: Colombia Applicable trade region(s): CAN (Andean community) Application to the chosen product: 1. Describe in a much details possible the specifications...

Words: 456 - Pages: 2

Free Essay

My Test

...| | | | | | | |INSPECTION / |STANDARD (1) |PROCEDURE AND/OR ACCEPTANCE CRITERIA |DOC |SAMPLE |CUSTOMER | |TEST ACTIVITY | | |Y/N(2) |SIZE |INSPECTION | |OPERATION ACTIVITY: 0010 |IN PROCESS | | | |VERIFY PART(S) HAVE RAW MATERIAL TRACEABILITY IS AVAILABLE BEFORE USE | | | | |DRSV010 |CAMERON STANDARD |METHOD- QUALIFIED PERSONNEL TO VERIFY APPLICABLE MATERIAL IS CORRECT BEFORE USE |YES |100% |- | |RAW MATERIAL | |STANDARD- CAMERON STANDARD | | ...

Words: 475 - Pages: 2

Premium Essay

Calibration of a Chromel-Alumel Thermocouple

...| Calibration of a Chromel-Alumel Thermocouple | | Module Title: Engineering ApplicationsModule Tutor: | Student Name: | Submission Date: | | Contents Objectives 3 Introduction 3 Apparatus 3 Theory 3 Results 4 Discussion 4 Conclusion 6 References 6 Calibration of a Chromel-Alumel Thermocouple Objectives * Use a thermometer to tell the changing temperatures of water * To become familiar with the principles of temperature measurement Introduction The purpose of this experiment was to measure the change in voltage in boiling water (electromotive force) against the temperature (measured by a thermometer) when cold water was added at stages. The voltage readings were taken using a Chromel-Alumel thermocouple. 1 6 5 2 Apparatus 1. Immersion heater and power supple 2. 500ml beaker half full with water 3. 3 1000ml beaker full with water 4. Thermometer 5. Thermocouple 6. Digital multimeter 7. Clamp and stand 7 4 Theory Refer to lab sheet for method used. A thermocouple is thermoelectric temperature sensor (which means it can measure temperature using electricity). A thermocouple consists of two wires that are made of different metals to one another which are coupled at the measurement junction and extended to the reference junction. [1] [1] A thermometer is “an instrument for measuring and indicating temperature, typically one consisting of a narrow, hermetically sealed glass tube marked...

Words: 957 - Pages: 4

Premium Essay

Analytical Chemistry Thesis

...Method Development For Analysis Of Pesticides In Water Using Chromatography And Mass Spectrometry An Abstract of a Thesis Presented to the Faculty of Chemistry Department Western Illinois University In Partial Fulfillment of the Requirements for the Degree Master of Science By Prashanth Goud Kunuru September 2013 This thesis follows the style and format of the Journal of Analytical Chemistry ABSTRACT Pesticides are chemicals, artificial or natural substances, used for pest control. They have negative influence on human health when accumulated in the body. Numerous studies have clearly linked pesticides exposures to various diseases including cancer and Parkinson’s disease. Water bodies often serve as a key receiver of harmful pollutants emitted from industrial and domestic sources. Pesticides can often enter water through sewage plants, sewer overflows, surface water runoffs, and leaching from agricultural applications or even through improper disposal. Regulatory issues and public concerns over pesticide residues in water supplies have been increasing, and it is important to develop a reliable method for routine test of pesticides in water supplies. Presented here is a novel solid phase extraction technique using disposable pipette extraction (DPX) followed by chromatography and mass spectrometry analysis for determination of pesticides in water. The purpose of the current study was to evaluate the suitability of high pressure...

Words: 7995 - Pages: 32

Free Essay

Spychology

...Experimental Physiology (1986) 71, 467-473 Printed in Great Britain FREE INTRACELLULAR MAGNESIUM CONCENTRATION IN FERRET VENTRICULAR MUSCLE MEASURED WITH ION SELECTIVE MICRO-ELECTRODES L. A. BLATTER AND JOHN A. S. McGUIGAN Department of Physiology, University of Berne, Buehlplatz 5, 3012 Berne, Switzerland (RECEIVED FOR PUBLICATION 2 AUGUST 1985) SUMMARY The free Mg concentration ([Mg]i) was measured in ferret ventricular muscle using singlebarrelled Mg-sensitive micro-electrodes. The electrodes were calibrated in solutions containing the [K]i and [Na]i measured in this tissue. In thirteen measurements from seven experiments conforming to strict criteria for calibration and penetration the estimated mean [Mg]i was 04 mmol/l. This is to be regarded as an upper limit since it was estimated by linear extrapolation of the calibration curve. INTRODUCTION Ionized Mg plays an essential role in the regulation of a large number of cellular functions. It is a co-factor in many enzymatic reactions, influences tension development in muscle and modulates amongst others the ionic transport systems in mitochondria (see reviews by Gunther, 1977; Shine, 1979; and Ackerman & Nicholls, 1983). Despite its importance in cellular mechanisms, estimations of the free Mg ([Mg]i) by various methods range from 0- 1 mmol/l to 6 mmol/l (reviews: Flatman, 1984; Gupta, Gupta & Moore, 1984) mainly because of the lack of a simple direct method for its measurement. Direct measurement of [Mg]i is now possible using...

Words: 3199 - Pages: 13

Premium Essay

Nt1310 Unit 1 Lab Report

...supply and multimeter measurements were recorded as meter data for the BI, 5V, and 5I measurements. The EyasSAT GUI was used to measure and record the “raw counts” at the BI, 5V, and 5I locations. The “raw counts” measurements were recorded as counts data. The EyasSAT GUI was used to measure and record the current at the 5I and BI locations as well as measure and record the voltage at the 5V location. The current and voltage measurements made by the EyasSAT GUI were recorded as data collected before calibration. This process was done for five different values of load resistors. The load resistor was removed and the DC power supply voltage was measured and recorded as meter data. The EyasSAT GUI was used to measure and record the “raw counts” for the BV location. The “raw counts” data was recorded as counts data. The EyasSAT GUI was used to measure and record the voltage at the BV location. The voltage measurement made by the EyasSAT GUI was recorded as data collected before calibration. This process was repeated for DC power supply outputs of 10, 9, 8, 7, and 6 volts.   A linear regression and coefficient of determination was calculated for the meter data and the counts data for each of the four measurements, BV, BI, 5V and 5I. The counts data was put on the x-axis and the meter data was put on the y-axis. Equations (1) – (3) were used to calculate the linear regressions and coefficients of determination for each data set. Equation (1) shows how the slope of the line (b) was calculated...

Words: 1717 - Pages: 7

Free Essay

Inertia Bending of a Connecting Rod Experiment

...moment-----------------------------------------------------------------------------2 1.3.2 The Right Angle Assumption--------------------------------------------------------------------------5 2. Apparatus----------------------------------------------------------------------------------------------6 * 3. Data-----------------------------------------------------------------------------------------------------7 * 4. Procedur-----------------------------------------------------------------------------------------------7 * 4.1 Recording-------------------------------------------------------------------------------------------7 * 4.2 Calibration------------------------------------------------------------------------------------------7 5. Results--------------------------------------------------------------------------------------------------8 * 5.1 Determination of the calibration constant----------------------------------------------------8 5.2 Calculation of the Bending Moments----------------------------------------------------------9 5.2.1 Analysis of the oscilloscope traces----------------------------------------------------------9 5.2.2 Calculation of...

Words: 2873 - Pages: 12

Premium Essay

Supply

...>>> INSTRUMENTATION KEY FEATURES DESIGN ■ Compact and rugged ■ One button operation ■ Easy to read front mounted backlit display ■ Over-molded boot OPTIONAL SLIP-ON PUMP ■ ■ SENSORS AND CALIBRATION ■ Three sensors. O2, LEL and CO, H2S, or combined CO/H2S sensor ■ User replaceable sensors ■ One button auto calibration ■ Calibration due reminder ■ Sensor span reserve indicators during calibration for predictive maintenance BATTERIES ■ Software controlled pump Powered by the Biosystems MultiPro battery ■ Automatic low flow alarm ■ Automatic leak test before every use RECORDING ■ Automatic event logger (20 events, including sensor type, max, average, time and duration) ■ Black-Box recorder with over 40 hours of storage ■ Available with optional datalogging for convenient and easy downloading to a computer Interchangeable alkaline or rechargeable (Li-Ion) versions ■ Up to 22 hours run time depending on battery type ■ Drop-in charger ■ Recharges in less than 6 hours BIOSYSTEMS MULTIPRO The Biosystems MultiPro is the smallest and easiest to use confined space gas detector offered by Sperian Instrumentation. Extremely affordable, it is packed with features. It offers real time readings of up to 4 gases simultaneously, one-button operation including calibration, an easy to read LCD display, audible, visible and optional vibrating alarms, standard event logger and black box datalogger, an IrDA port for easy communication with a PC, and an optional sample draw...

Words: 1050 - Pages: 5

Premium Essay

Mama

...1. Introduction 1 1.1. About ISO 9000 1 1.2. Purpose of internal quality audit 1 1.3. Report structure 1 2. Audit review 2 2.1. Non-compliances 2 2.1.1. Case study 2 – Service and parts processes 2 2.1.2. Case study 3 – QMS manual 3 2.2. Potential for improvement 4 2.2.1. Case study 2 – Service and parts processes 4 2.2.2. Case study 3 – QMS manual 5 2.3. Good practice 5 2.3.1. Case study 2 – Service and parts processes 5 2.3.2. Case study 3 – QMS manual 6 3. Summary 8 4. Appendix 11 Introduction About ISO 9000 ISO 9000 is a family of international standards related to quality management systems (QMS) that facilitates mutual understanding across domestic and international trade. The standards are published by the International Organization for Standardization and available in the UK through the British Standard Institute. ISO 9000:2008 sets down the principles behind the quality management and specifies the QMS terminology, thus forms the basis for the standards within the ISO 9000 family. ISO 9001:2008 specifies the requirements for a QMS an organisation has to fulfil to meet the standard. ISO 9004:2008 provides guidelines for a QMS that aim to improve the performance of the organisation and the satisfaction of both stakeholders and customers. ISO 9000 represents an international consensus on good quality management practices and can be understood as a generic framework to manage a business’ processes to deliver superior quality...

Words: 3300 - Pages: 14