Pain Physician, Volume 4, Number 4, pp 343-348 2001, American Society of Interventional Pain Physicians® ISSN 1533-3159

Case Study

Radiation Exposure to a Physician Performing Flouroscopically Guided Caudal Epidural Steroid Injections
Kenneth P. Botwin, MD*, Eric D. Freeman, DO, Robert D. Gruber, DO, Francisco M. Torres-Ramos, MD, Constantine G. Bouchlas, MD, Joseph T. Sanelli, DO, and Ashraf F. Hanna, MD
This study was designed to investigate radiation exposure to a physician performing fluoroscopically guided caudal epidural steroid injections. The prospective study design included 100 consecutive fluoroscopically guided caudal epidural steroid injections performed on patients with radiculitis from either herniated nucleus pulposus or lumbar spinal stenosis. Radiation exposure was monitored with the assistance of a radiological technologist (RT) who allocated four dosimetry badges to all physicians performing fluoroscopically guided caudal epidural steroid injections on consecutive patients being treated for radicular pain. The badges were placed on the ring finger, glasses and both the inside and outside of the lead apron worn by the physician. In addition, the RTs also wore a marked badge outside his/her lead apron. A control badge was placed 67 inches away from the fluoroscopy table, and a second control badge was located in a desk over 500 feet away from the procedure, to monitor ambient radiation. The average fluoroscopy time per procedure was 12.55 seconds. The average/cumulative exposure per procedure was 4.10/410 mREM at the “ring” badge, 2.47/247 mREM at the “glasses” badge, 3.98 /398 mREM at the “outside apron” badge and 0.15/15 mREM at the “inside” apron; no radiation was detectable at the “outside room” control badge. The RT’s average exposure during these procedures was below the limit of detectability. . Radiation exposure to the physician needs to be considered and minimized in the performance of spinal interventional procedures. Our study demonstrates that radiation exposure to the physician performing fluoroscopically guided caudal epidural steroid injections is well within safety limits when he/she adheres to proper technique. Keywords: Caudal epidural steroid injections, fluoroscopy, radiation safety

Fluoroscopically guided caudal epidural steroid injections Incorrect needle placement in caudal epidural injections are commonly utilized in the treatment of bilateral lumbosacral radiculitis. The methodology and efficacy of this procedure have been reviewed elsewhere (1, 2, 3). The indications for caudal epidural nerve blocks are numerous. They have been used to treat benign pain syndromes such as diabetic polyneuropathy, post-herpetic neuralgia, complex regional pain, vertebral compression fractures and pelvic pain syndromes (1, 2, 4, 5, 6, 7, 8, 9). Other indications include surgical and obstetrical anesthesia and cancer-related pain (7). The caudal approach to the epidural space is helpful in patients who have had prior lumbar surgery, which makes the lumbar approaches more difficult (7). From Florida Spine Institute Clearwater, Florida. *Dr. Botwin is director of Florida Spine Institute. Address correspondence: Dr. Kenneth Botwin, 2250 Drew Street, Clearwater, FL 33765. occurs with relative frequency when performed without the use of fluoroscopic guidance, with a 12.5% to 38.3% incidence (10, 11, 12). Thus, with the advent of fluoroscopy, more procedures are being performed with this method to avoid incorrect needle placement. Fluoroscopy is used in many procedures, including swallowing studies; urologic evaluations; peripheral joint injections; and, perhaps most commonly, interventional spine procedures. Fluoroscopy is one of the two types of x-ray procedures, the other being radiography. Conventional fluoroscopy, which consists of an x-ray tube located above a fixed examining table, has been widely replaced by C-arm fluoroscopes with image intensification for use in spinal injection procedures. The C-arm permits the physician to rotate and angle the x-ray tube around the patient while he/ she rests on a radiolucent support table. Image intensification is achieved through the addition of an image inten-

343

Botwin et al • Radiation Exposure of Caudal Epidurals

344

sifier, which receives remnant x-ray beams that have passed through the patient and converts them into light energy, thereby increasing the brightness of the displayed image and making it easier to interpret. To our knowledge there have been no published studies to assess the radiation exposure to the physician performing fluoroscopically guided caudal epidural injections. Studies have been performed to evaluate the risks to orthopedic surgeons using fluoroscopy (13, 14), cardiologists performing cardiac catheterization (15), urologists doing endourologic procedures (16) and interventional radiologists performing a variety of visceral and peripheral angiographic procedures (17). The measurement for recording exposure is the rad equivalent man (REM) or milliequivalent man (mREM), which can be measured using a dosimetry badge. The reader is referred to other more extensive reviews of radiation quantities, units and terminology (17, 18). These studies and ours studied exposure levels using dosimetry badges, which allow the extent of exposure to be recorded. This study was performed to evaluate radiation exposure to a physician performing fluoroscopically guided caudal epidural steroid injections. Review of the current medical literature did not reveal any study on radiation exposure to the physician performing these procedures. This is a prospective, cohort design study of 100 fluoroscopically guided caudal epidural steroid injections performed in 100 patients with radicular pain from spinal stenosis to determine the radiation exposure to the physician performing these injections. METHODS One hundred consecutive patients presented to a multidisciplinary spine care practice with complaints of lower back and radicular pain. As part of a nonoperative treatment plan, patients received a combination of analgesics, anti-inflammatory medications, referral to physical therapy and fluoroscopically guided caudal epidural steroid injections. Inclusion criteria consisted of consecutive patients who presented to the spine care center with radicular pain from herniated nucleus pulposus or lumbar spinal stenosis, which was treated with fluoroscopically guided caudal epidural steroid injections. Exclusion criteria included allergy to iodine dye or any component of the injection, and a history of prior surgery.

Procedures were performed in an ambulatory surgical center by one of seven physicians experienced in fluoroscopically guided caudal epidural injections. Similar technique was used in all reviewed procedures. Patients were placed in the prone position on a radiology table. A wedge-shaped pillow was placed under the hips to tilt the pelvis and bring the sacral hiatus into greater prominence. The sacrococcygeal area was prepared using an iodine-based antiseptic solution (Povidone Iodine, USP solution, The Clinipad Corporation, Rocky Hill, CT), and an alcohol solution (Kendall Webcol Alcohol Prep 70% Isopropyl Alcohol, Marsfield, MA). The physician, using the sterile, gloved middle finger of the dominant hand, then localized the tip of the coccyx through palpation. In this position the area under the proximal interphalangeal joint was marked. Using a fluoroscope (Compact 7600, OEC, Salt Lake City, UT), a 22-gauge, 3.5-inch/ 90 mm spinal needle (Quincketype point, Luer lock, Spinocan, Becton Dickinson, Franklin Lakes, NJ) or 20-gauge, 3.5-inch/90 mm Tuohy Spinal needle (Becton Dickinson, Franklin Lakes, NJ) was guided under intermittent fluoroscopy to the midline of the sacral hiatus. A lateral fluoroscopic view was used to confirm that the needle was in the caudal epidural space. Aspirations were routinely performed. If negative for aspirate, Isovue M-300 (Iopamidol Injection, Bracco Diagnostics, Princeton, N.J.) 1 mL was instilled to confirm epidural flow of the injectate and to rule out intravascular, intrathecal and/or soft tissue infiltration. Once an epidurogram was obtained, a solution of 13 mL of 0.5% preservative-free Xylocaine (Lidocaine HCL Injection, Astra pharmaceuticals, Westborough, MA) and 12 mg betamethasone acetate (Celestone® Soluspan®, Schering, Kenilworth, NJ) was injected. Plain radiographs in the anteroposterior and lateral views had been taken after all injections to document both the contrast pattern and needle placement. All patients were monitored by pulse oximetry, blood pressure and ECG prior to, during, and after the procedure. Patients were transferred to the recovery unit for 40 minutes. All patients were seen by the physician who performed the injection and by a registered nurse prior to discharge. All epidural injections were performed at an ambulatory care center using C-arm fluoroscopic guidance. The Carm fluoroscope (OEC Compact 7600) used was operated in snapshot (pulsed-imaging) mode. X-ray voltages are measured in kilovolts peak (kVP). In the pulsed mode, a high-energy penetrating x-ray beam was created with the

Pain Physician Vol. 4, No. 4, 2001

Botwin et al • Radiation Exposure of Caudal Epidurals

345

image intensifier delivering a high kVp (The peak value of high-voltage generator output in the interval after a 20millisecond delay period to the end of the exposure) was based on the patient’s weight and a fixed current of 8 mA using the snapshot mode. In this mode beam collimation constricted the x-ray beam to the lumbar spine. Radiation exposure was monitored using a Landauer (Glenwood, IL) dosimetry badge with a lower limit of detectability of 1 mREM. One radiographic technologist (RT) allocated four badges to physicians prior to the procedures. The badges were clearly marked as (1) outside, (2) inside, (3) ring, and (4) glasses. The “outside” badge was placed outside the lead apron worn by all the physicians, which was of 0.55 mm thickness (ProTech, Palm Beach Gardens, FL). The “inside” badge was placed at the shirt pocket level under the apron. The “ring” badge was placed on the ring finger of the dominant hand. The “glasses” badge was placed on the frame of lead glasses (ProTech leaded eyewear, Palm Beach Gardens, FL) that the physicians wore during the procedure. In addition, the RTs also wore a clearly marked badge outside their lead apron. A control badge was arbitrarily placed on the wall 67 inches away horizontally and 56 inches vertically, from the fluoroscopic table in the treatment room. A second control badge was also placed over 500 feet outside the fluoroscopy room to evaluate for any ambient radiation. When badges were not in use, they were all placed together with the second control badge, outside the treatment room. All RTs stood over 6 feet away from the fluoroscope. The RTs also entered into a daily log the patient’s name, date of procedure, medical record number procedure level, total fluoroscopy time, voltage in kVp, amperage in milliamps, physician performing procedure and weight of the patient. Following all procedures the dosimetry badges were sent to Landauer for interpretation by the RT. The radiation dosimetry report from Landauer included the dose equivalent mREM (the unit of occupational radiation exposure used to monitor personnel exposure devices such as film badges) for the period in which the injections were performed. Separate readings were obtained for all badges for deep-dose, eye and shallow-dose equivalents in mREM. Deep-dose equivalent applies to external whole-body exposure and is the dose equivalent at a tissue depth of 1 cm. Eye-dose equivalent applies to the eternal exposure of the lens of the eye and is taken as the dose equivalent at a tissue depth of 0.3 cm. Shallow-dose equivalent applies to the external exposure of the skin or an extremity, and is taken as the dose equivalent at a tissue depth of 0.007 cm

Table 1. Patient data
Total number patients Total fluoroscopy time (seconds) Range fluoroscopy time (seconds) Mean age (years) Mean weight (pounds), men Mean weight (pounds), women 100 1255 2- 33 6 5. 7 2 199.4 178.5

averaged over an area of 1 sq cm. The total amount of mREM for each badge was obtained, and the amount of mREM per second of fluoroscopy use was calculated. All reports were reviewed and compiled by the radiographic technologist. RESULTS A total of 100 procedures on 100 patients (47 women and 53 men) was performed. All patients had radicular pain from lumbar spinal stenosis or herniated nucleus pulposus. The total fluoroscopy time of all 100 procedures was 125.5 seconds. The mean weight of the patients studied was 189.37 lbs. The weight range of men was 140 to 277 lbs, and the weight range of women was 110 to 379 lbs. The mean male patient weighed 199.4 lbs, and the mean female patient weighed 178.5 lbs. The mean age of patients was 65.72 years. The mean fluoroscopy time was 12.55 seconds. The mean kVp received by women patients was 93.375 at 8mA and 91.0769 at 8mA in male patients. (See Table 1). The exposure to the physician was calculated in mREM for all 100 procedures, from dosimetry badges. The exposure at the “outside” apron was 398 mREM (deep-dose equivalent) and at the “inside” apron was 15 mREM; “glasses” exposure was 247 mREM (eye-dose equivalent)

Table 2. Mean mREM per procedure in all 100 patients with an average fluoroscopy time of 12.55 seconds
Interventionalist, "ring" Interventionalist, "glasses" Interventionalist "outside" apron Room control (inside room) 4.10 2.47 3.98 0.22

Pain Physician Vol. 4, No. 4, 2001

Botwin et al • Radiation Exposure of Caudal Epidurals

346

and “ring” exposure was measured to be 410 mREM (shallow-dose equivalent). The control badge which was placed inside the room 67 inches away horizontally and 56 inches vertically from the fluoroscopy table recorded an exposure of 22 mREM deep-dose equivalent. The exposure recorded from both the outside and inside dosimetry badges on the technicians and the outside room badge was 0 mREM deep-dose equivalent. We were able to calculate the exposure per procedure using the average procedure time of 12.55 seconds to find an “outside” badge deep-dose equivalent of 3.98 mREM, a “glasses” eye-dose equivalent of 2.47 mREM a “ring” shallow-dose equivalent of 4.10 mREM and an in-room control, deep-dose equivalent of 0.22 mREM (See Table 2). DISCUSSION The use of epidural steroid injections in the treatment of lumbosacral radiculopathy has become an important procedure that is both diagnostic and therapeutic. The caudal epidural injection is one of several injection techniques in which the lumbar epidural space can be accessed. The presumed ease and accessibility of the epidural space have made caudal epidural injections a favorite in specific patient populations. Cathelin first described the procedure in 1901 (19). The introduction of midline lumbar epidural technique by Pages in 1921 caused the caudal approach to fall out of favor (20). The caudal approach was re-introduced as a technique for pain relief in childbirth by Hingson and Edwards in 1943 (21). Epidural steroid injections have been used in the treatment of lumbar radicular pain syndromes since 1952 (22-23). They were first reported in the United States in 1960 to benefit conditions causing nerve root irritation (23, 24). These injections were performed “blind” (without fluoroscopic guidance) using a caudal technique (24).

Radiation risks to the physician and assisting personnel are evaluated using the maximum safe allowable exposure limits which have been established by the National Council on Radiation Protection and Measuring as a maximum permissible dose (25). The recommendations for occupational radiation exposure can be seen in Table 3. The current estimation of risk from radiographic exposure to a specific body part is based on the biologic effects of whole-body exposure converted by weight factors specific for individual organs and tissues. This concept was adopted by the International Commission on Radiological Protections in 1977 and was modified in 1991. Termed the effective dose equivalent, the calculation was adopted by authoritative bodies that determine radiation risks and recommend protective measures (See Table 4).

Table 4. Specific organ cancer risks of radiation (per 10,000 per Sv or per 1,000,000 per REM)
Radiation organ or cance r Breast Thyroid Lung Leukemia Stomach Brain Colon Liver Lymphoma Uterus Salivary Glands Ovary Bladder Bone Esophagus Pancreas Paranasal sinuses Probability of induce d cance r 50- 200 50- 150 50 15 - 2 5 10- 20 5- 20 10- 15 10- 15 4 - 12 7- 10 5- 10 8 4- 7 2- 5 2- 5 2- 5 2- 5

Table 3. General radiation exposure guides
Type of Expos ure Whole body (head and trunk) Lens of eye Extremities Skin
Modified from Reference 25.

Annual Limit 5 REM/year. total effective dose equivalent 15 REM/year 50 REM/year 50 REM/year

From International Commission on Radiological Protection: Recommendations of the International Commission on Radiation Protections 26 (30). * 100 REM = 1 Sievert (Sv); 1/100 mREM = 1 mSv.

Pain Physician Vol. 4, No. 4, 2001

Botwin et al • Radiation Exposure of Caudal Epidurals

347

When evaluating our data, it can be seen that the radiation exposure appears to be well within these regulations. If we extrapolate our data to 1000 fluoroscopically guided caudal epidural steroid injections which could be performed annually per clinician at our center, then a total radiation exposure of 4.10 REM at the “ring” badge, 2.44 REM at the “glasses” badge and 3.88 REM at the “outside apron” badge all would fall within limits. However, depending on the volume of procedures in the physicians clinic the exposure will increase accordingly and must be followed closely. The absolute exposure to the physician must be calculated on an individual basis and must take into account that the radiation exposure will be cumulative over a lifetime. With this in mind, leaded aprons, glasses, thyroid shields and lead barriers seem most appropriate to minimize the physician’s exposure. This has also been found by other authors (26). It should be noted the long-term effects of low-dose radiation are uncertain. The exposure in our study was greatest to the hands, then eyes, of our physicians, which may argue for the use of leaded gloves in addition to leaded glasses. Also, extended tubing for administering radiographic contrast may be of benefit. Radiation exposure results from the primary beam and scatter radiation reflected off the floor and the patient. The cardinal principles of radiation protection are: (1) maximize the distance from the radiation source; (2) use shielding materials; and (3) minimize exposure time. Technical factors using the fluoroscope may serve to reduce both occupational and patient exposure. Using pulsed imaging during fluoroscopy has been shown to reduce overall exposure by 20% to 75% (27, 28). The usage of collimation also reduces unnecessary radiation exposure (28). A radiation safety program must serve to minimize exposure to ionizing radiation while promoting safe and effective use of radiation sources in diagnosis, therapy and research. One goal of such a program is to keep the risks to radiation workers to levels that are comparable to those of other safe occupations. The pervading philosophy is that of “as low as reasonably achievable” (29). CONCLUSION Adhering to simple rules of radiation safety can minimize the cumulative exposure to the spinal interventionalist. These include measuring the distance between the interventionalist and the source of radiation; decreasing the overall time of exposure; and shielding susceptible

areas with leaded aprons, thyroid shields, leaded glasses and leaded gloves, as well as being proficient in guiding needles under the fluoroscope. Our study demonstrates that the radiation exposure to the physician during the fluoroscopically guided caudal epidural steroid injections is well within safety guidelines. ACKNOWLEDGMENTS The Authors wish to thank Nicole Belsanti, RRT, Diane Hodges, RRT, and Carol Barragan, Medical Secretary for their assistance with this manuscript. REFERENCES
1. Gardner WJ, Goebert HW, Sehgal AD. Intraspinal corticosteroids in the treatment of sciatica. Trans Am Neurol Assoc. 1961;86:215. Yates DW. A comparison of the types of epidural injections commonly used in the treatment of low back pain and sciatica. Rheum Rehab 1978;17:181-186. Sharma RK. Indications, techniques and results of caudal epidural injections for lumbar disc retropulsion. Postgrad Med J 1977; 53:1-6. Bush K, Hillier S. A controlled study of caudal epidural injections of triamcinolone plus procaine for the management of intractable sciatica. Spine 1991; 16:572-575. Mount HTR. Epidural injection of hydrocortisone for the management of the acute lumbar disc protrusion. In Morley TP (ed). Current Controversies in Neurosurgery. Philadelphia, Saunders, 1976, pp 67-72. Goebert HW, Jallo SJ, Gardner WJ, et al. Painful radiculopathy treated with epidural injections of procaine and hydrocortisone acetate: Results in 113 patients. Anesth Analg 1961;140:130-134. Willis RJ. Caudal epidural blockage. In Cosins MJ, Brindenbaugh DO (eds). Neural Blockage. Philadelphia, JB Lippincott, 1988, pp 376-377. Waldman SD. Reflex sympathetic dystrophy. Intern Med 1990; 11:62068. Waldman SD. Acute herpes zoster and postherpetic neuralgia. Intern Med. 1990; 11:33-38. White AH, Derby R, Wynn G. Epidural injections for the treatment of low back pain. Spine 1980; 5:78-86. Renfrew DL, Moore TE, Kathol MH. Correct placement of epidural steroid injections: Fluoroscopic guidance and contrast administration. AM J Neuroradiology 1981; 12:1003-1007, 1981. Sitz and Somer. Accuracy of “blind” verses fluoroscopically guided caudal epidural injections. Spine 1999; 24:1371-1376. Giachino AA, Cheny M. Irradiation of the surgeon during pinning of femoral fractures. JBJS 1980; 62:227229.

2.

3.

4.

5.

6.

7.

8. 9. 10. 11.

12.

13.

Pain Physician Vol. 4, No. 4, 2001

Botwin et al • Radiation Exposure of Caudal Epidurals

348

14.

15.

16.

17.

18. 19.

20. 21.

22.

Miller ME, Davis ML, MacClean CR et al. Radiation exposure and associated risks to operating room personnel during use of fluoroscopic guidance for selected orthopedic surgical procedures. JBJS 1983; 65-A:1-4. Miller, SW, Castrovovo FP. Radiation exposure and protection in cardiac catheterization Laboratories. AJ Cardiology 1985; 55:171-176. Giblin JG, Rubenstein J, Taylor A et al. Radiation risk to the urologist during endourologic procedures, and a new shield that reduces exposure. Urology 1996; 48:624-627. Gruber RD, Botwin KB, Shah CP. Radiation Safety for the Physician. Pain Procedures in Clinical Practice, ed. 2, Hanley and Belfus 2000, p 25. Wycoff HO. The international system of units. Radiology 1978; 128:833-835. Cathelin MF. Une nouvelle voie d’injection rachidienne. Methode du injections epidurales pas le procedue du canal sacre. D R Soc Biol 1901; Par 53:452. Pages E. Anestesia metamerica. Rev Sanid Mil Madr 1921; 11:351-380. Hingson RA, Edwards WB. An analysis of the first ten thousand confinements managed with continuous causal analgesia with a report of the authors’ first one thousand cases. JAMA 1943; 125:538. Robechhi A, Capra R. L’idrocotisone (composto F), Rime esperinze cliniche in campo reumatologico,

23.

24. 25.

26.

27. 28.

29.

30.

Minerva Med 1952; 98:1259-1263. Benzon H. Epidural steroid injections for low back pain and lumbosacral radiculopathy. Pain 1986; 24: 277-295. Brown JH. Pressure caudal anesthesia and back manipulation. Northwest Med 1960; 59:905-909. National Council on Radiation Protection and Measurements. Ionizing Radiation Exposures of the Population of the United States. Report No. 116, Washington, DC, NCRP, 1993. Marx VM, Niklason L, Mauge EA. Occupational radiation exposure to the interventional radiologist: A prospective study: JVIR 1993; 3:597-606. Adelstein SJ. Uncertainty and relative risks of radiation exposure. JAMA 1987; 258:655-657. Hernandez RJ, Goodsitt MM. Reduction of radiation dose in pediatric patients using pulsed fluoroscopy. Am J Roentgerol 1996;167:1247-1253. National Council on Radiation Protection and Measurements. Implementation of the Principle of as Low as Reasonable Achievable (ALARA) for Medical and Dental Personnel. NCRP Report no. 107. Bethesda, MD, NCRP, 1990. International Commission on Radiological Protection. Recommendation of the International Commission on Radiation Protection 26. Ann Int Commission Radiat Prot 1977; 1:1-53.

Pain Physician Vol. 4, No. 4, 2001

Similar Documents

Premium Essay

Hsa 525

...Introduction Brandywine Homecare, a not-for-profit business, had revenues of $12 million in 2007. Expenses other than depreciation totaled 75 percent of revenues, and depreciation expense was $1.5 million. All revenues were collected in cash during the year and all expenses other than depreciation were paid in cash. My report will contain the answers to the following questions: 1. Construct Brandywine’s 2007 income statement. 2. What were Brandywine’s 2007 net income, total profit margin, and cash flow? 3. Suppose the company changed its depreciation calculation procedures (still within GAAP) such that its depreciation expense doubled. How would this change affect Brandywine’s net income, total profit margin, and cash flow? 4. Explain the difference between cash and accrual accounting. Be sure to include a discussion of the revenue recognition and matching principles. 5. Explain the difference between equity section of a not-for-profit business and an investor-owned business. 1. Construct Brandywine’s 2007 income statement. |Brandywine Homecare | | | |Income Statement | | | |Year Ended December 31, 2007 | | ...

Words: 457 - Pages: 2

Free Essay

Hsa 525

...Financial Statement Analysis In this paper, the writer will perform an analysis of Universal Health Services (UHS) financial statements. The organization has a group arrangement, in which they are the parent corporation of numerous subsidiaries. The company focuses on acute care hospitals, behavioral health centers, surgical hospitals, ambulatory surgery centers, and radiation oncology centers. (Universal Health Services, 2015) Universal Health Services, a New York Stock Exchange company, has its headquarters located in Pennsylvania but has several locations throughout the United States including Puerto Rico, U.S Virgin Islands, and the United Kingdom. (Universal Health Services, 2015) Financial Well Being of Universal Health Services The health care industry can be a lucrative business, for UHS it is seen in their annual report provided to stakeholders. The report informs stakeholders what the company has done to expand its financial standing in order to grow within the year. UHS reached many goals within their past fiscal year that ended in December of 2014 both financially and as a corporation. First their net revenue reached 8.07 billion an all time high for the organization’s 35 years of business compared to last year. Looking at the adjusted net income that is attributable to UHS it increased from $452.1 million in 2013 to $581.8 million in 2014 which led to a 29 percent increase. The company’s stock price valued 37 percent, closing 2014 at $111.26 per share. In...

Words: 979 - Pages: 4

Premium Essay

Hsa/525

...Using Financial Ratios Evangeline Hampton HAS/525 Debra Beazley May 9, 2014 Using Financial Ratios Introduction Suggest the financial ratio that most financial analysts would use to evaluate the financial condition of the company. Provide support for your rationale. Financial ratios are tools utilized to examine the financial condition and performance of company by financial analyst. Financial analysts look at many different ratios. The most important ratios that are used to analyze the financial health of a company are liquidity and profitability ratios. Financial ratios are important because they allow financial managers to evaluate opportunities. Analysts study a company’s financial statements and are particularly concern with return on investment in the various assets of the company and in the efficiency of asset management. Getting the ratios numbers involves analysis and use of the financial statements of a firm. These statements attempt many things. First, the statements reveal the assets and liabilities of a business firm at a moment in time usually at the end of a year. Analyst can use these statements to compare present ratios with past ratios of the same company by using past statements. Analyst can also use ratios of other firms and compare them with firms of similar industry. Evangeline opinion is the financial analyst first look at profitability ratios, used to measure a company’s ability to generate earnings relative to its expenses and other costs...

Words: 294 - Pages: 2

Premium Essay

Hsa 525 Week 6

...HSA 525 WEEK 6 A+ Graded Tutorial Available At: http://hwsoloutions.com/?product=hsa-525-week-6 Visit Our website: http://hwsoloutions.com/ Product Description HSA 525 Week 6, SOLUTION 14-2, 14-3 Solutions 13-3,13-4.1,13-4.2 HSA 525 Week_6_Assignment 2 Financial Ratio The financial ratios are used for the purpose of determining the financial performance of an organization. There are various financial ratios which are taken into consideration by the financial analysts. The most important ratio which shall be taken into consideration is the net profit ratio. The net profit ratio of an organization is a particular ratio which is determined by the organization by deducting the expenses incurred by it from the revenues generated by it. If the net profit ratio of an organization is lower then, it can be said that, the organization does not have the ability or control over its expenses to a larger level. The organization compares its net profit with the competitors. The organization which is higher net profit ratio is considered to be better in comparison to the competitors of the organization (Shim & Siegel, 2007). There are various organizations which earn a lot of revenue but, if an organization will not have control over its expenses then, there will be the situation that, the organization will not be in the position to have profitability. The success of an organization is judged by its profitability other than anything else. Meeting Financial Obligations The...

Words: 426 - Pages: 2

Premium Essay

Hsa 525 Week 4

...HSA 525 WEEK 4 A+ Graded Tutorial Available At: http://hwsoloutions.com/?product=hsa-525-week-4 Visit Our website: http://hwsoloutions.com/ Product Description PRODUCT DESCRIPTION HSA 525 Week 4, week 4 analysis preview Introduction The organization Universal Health Services is a major health care organization in the United States of America. The organization provides various health care services. This paper will discuss financial health care of Universal Health Services. The strategy to improve the financial health of the organization will also be taken into consideration in the present situation. The financial analysis of the organization will be carried out for a period of three previous years. Review of the Financial Statements The financial statements of the organization will be used by various stakeholders. The stakeholders who will use the financial statements of the organization will include employees, investors, shareholders and the management of the organization (Shim & Siegel, 2007). In relation to the financial statements of the organization certain things are required to be taken into consideration. The revenues generated by the organization, the profits earned by the organization and the level of current assets and current liabilities of the organization are the crucial things which are necessary for the purpose of above mentioned stakeholders. In relation to the review of financial performance of the organization, it is provided that, the revenues...

Words: 539 - Pages: 3

Premium Essay

Brandywine Hsa 525

...Brandywine Homecare, a not-for-profit business, had revenues of $12 million in 2007. Expenses other than depreciation totaled 75 percent of revenues, which is $9 million, and depreciation expense was $1.5 million. Income statement is the one of the three financial statements. The other two are the balance sheet and the statement of the cash flows. Brandywine Homecare’s total profit margin of 12.5 percent shows that the homecare makes 12.5 cents on every dollar of total revenues. If the company doubled its depreciation, both net income and profit margin would be zero. The cash flow will be the same for both cases, which is $3 million. Cash accounting and accrual accounting are two similar methods of maintaining accurate accounting records. Accrual accounting is considered to be the standard accounting practice for most companies. Equity is the ownership claim against total assets. For investor-owned businesses, equity is the amount of owner-supplied financing. Equity for not-for-profit businesses is the amount of capital supplied by the, charitable contributions or other organization. Brandywine Homecare Brandywine Homecare, a not-for-profit business, has reported revenues of $12 million in 2007. Expenses other than depreciation totaled 75 percent of revenues, and depreciation expense was $1.5 million. It is reported that all revenues were collected in cash during the year and all expenses other than depreciation were paid in cash. Brandywine Homecare’s 2007 Income Statement ...

Words: 312 - Pages: 2

Premium Essay

Hsa 525 Week 2 Assignment

...HSA 525 WEEK 2 ASSIGNMENT A+ Graded Tutorial Available At: http://hwsoloutions.com/?product=hsa-525-week-2-assignment Visit Our website: http://hwsoloutions.com/ Product Description HSA 525 Week 2 Assignment, Solution to Assignment Exercise 4–1 Contracted Contractual Payer Full Rate Rate Allowance FHP $72.00 – $35.70 = $36.30 HPHP 72.00 – 58.85 = 13.15 Solution to Assignment Exercise 4–2 Other Managed Public Commercial Care Medicare Medicaid Programs Patients Insurance Contracts (1) Intensive Care Unit X (2) Laboratory X One suggested solution is as follows. Physical/ Cardiac/ Occupational Pulmonary Therapy Rehab Training Administrative Nursing Salaries X X Physical Therapist Salaries X Occupational Therapist Salaries X Solution to Assignment Exercise 5–2 Yes I believe the expense grouping………… HSA 525 Week 2 Assignment, Solution to Assignment Exercise 4–1 Contracted Contractual Payer Full Rate Rate Allowance FHP $72.00 – $35.70 = $36.30 HPHP 72.00 – 58.85 = 13.15 Solution to Assignment Exercise 4–2 Other Managed Public Commercial Care Medicare Medicaid Programs Patients Insurance Contracts (1) Intensive Care Unit X (2) Laboratory X One suggested solution is as follows. Physical/ Cardiac/ Occupational Pulmonary Therapy Rehab Training Administrative Nursing Salaries X X Physical Therapist Salaries X Occupational Therapist Salaries X Solution to Assignment Exercise 5–2 Yes I believe the expense grouping………………………………………. HSA 525...

Words: 417 - Pages: 2

Premium Essay

Hsa 525 Week 3 Assignment

...HSA 525 WEEK 3 ASSIGNMENT A+ Graded Tutorial Available At: http://hwsoloutions.com/?product=hsa-525-week-3-assignment Visit Our website: http://hwsoloutions.com/ Product Description HSA 525 Week 3 Assignment, Solution 6.1 Allocation of Indirect Costs to Radiology Departments CC#557 CC#558 CC#559 CC#560 CC#561 Indirect Cost Centers Indirect Costs Allocation Basis Diagnostic Radiology Ultrasound Nuclear Medicine CT Scan Radiation Therapy Total Transporters $550,000 A $110,000 $132,000 $88,000 $154,000 $66,000 $550,000 Receptionists $360,000 B $60,000 $36,000 $72,000 $108,000 $84,000 $360,000 File Room Clerks $117,000 C $90,000 $3,375 $13,500 $4,500 $5,625 $117,000 Managers $240,000 B $40,000 $24,000 $48,000 $72,000 $56,000 $240,000 Solution 7.1 Analyzing Mixed Costs 1 Calculation of Variable and Fixed Costs through high low method: Highest-Volume (in September) 1100 Total Cost at Highest Volume $7,150 Lowest-Volume (in August) 100 Total Cost at Lowest Volume $1,010 Variable Cost = (High Vol.-Low Vol.)/(Total Cost at High-Total Cost at Low) $6.14 per pack Solution 7.2 1 Calculating the Contribution Margin: Revenue $1,210,000 Variable Expenses $205,000 Fixed Expenses $1,100,000 HSA 525 Week 3 Assignment Solution 6.1 6.2 7.1 7.2 HSA 525 Week 3 Assignment, Solution 6.1 Allocation of Indirect Costs to Radiology Departments CC#557 CC#558 CC#559 CC#560 CC#561 Indirect Cost Centers Indirect Costs Allocation Basis Diagnostic Radiology Ultrasound...

Words: 432 - Pages: 2

Premium Essay

Hsa 525 Week 3 Assignment

...A++PAPER;http://www.homeworkproviders.com/shop/hsa-525-week-3-assignment/ HSA 525 WEEK 3 ASSIGNMENT HSA 525 Week 3 Assignment, Solution 6.1 Allocation of Indirect Costs to Radiology Departments CC#557 CC#558 CC#559 CC#560 CC#561 Indirect Cost Centers Indirect Costs Allocation Basis Diagnostic Radiology Ultrasound Nuclear Medicine CT Scan Radiation Therapy Total Transporters $550,000 A $110,000 $132,000 $88,000 $154,000 $66,000 $550,000 Receptionists $360,000 B $60,000 $36,000 $72,000 $108,000 $84,000 $360,000 File Room Clerks $117,000 C $90,000 $3,375 $13,500 $4,500 $5,625 $117,000 Managers $240,000 B $40,000 $24,000 $48,000 $72,000 $56,000 $240,000 Solution 7.1 Analyzing Mixed Costs 1 Calculation of Variable and Fixed Costs through high low method: Highest-Volume (in September) 1100 Total Cost at Highest Volume $7,150 Lowest-Volume (in August) 100 Total Cost at Lowest Volume $1,010 Variable Cost = (High Vol.-Low Vol.)/(Total Cost at High-Total Cost at Low) $6.14 per pack Solution 7.2 1 Calculating the Contribution Margin: Revenue $1,210,000 Variable Expenses $205,000 Fixed Expenses $1,100,000 HSA 525 Week 3 Assignment Solution 6.1 6.2 7.1 7.2 HSA 525 Week 3 Assignment, Solution 6.1 Allocation of Indirect Costs to Radiology Departments CC#557 CC#558 CC#559 CC#560 CC#561 Indirect Cost Centers Indirect Costs Allocation Basis Diagnostic Radiology Ultrasound Nuclear Medicine CT Scan Radiation...

Words: 421 - Pages: 2

Premium Essay

Hsa 525 Homework Week 9

...HSA 525 Homework Week 9 16.1 - On a typical day, Park Place Clinic writes $1,000 in checks. It generally takes four days for those checks to clear. Each day the clinic typically receives $1,000 in checks that take three days to clear. What is the clinics average net float? 16.2 - Drugs 'R Us operates a mail order pharmaceutical business on the West Coast. The firm receives an average of $325,000 in payments per day. On average, it takes four days for the firm to receive payment, from the time customers mail their checks to the time the firm receives and processes them. A lockbox system that consists of 10 local depository banks and a concentration bank in San Francisco would cost $6,500 per month. Under this system, customers' checks would be received at the lockbox locations one day after they are mailed, and the daily total would be wired to the concentration bank at a cost of $9.75 each. Assume that the firm could earn 10 percent on marketable securities and that there are 260 working days and hence 260 transfers from each of the ten lockbox locations per year. (a) What is the total annual cost of operating the lockbox system? (b) What is the dollar benefit of the system to Drugs ‘R Us? (c) Should the firm initiate the lockbox system? 16.4 - Langley Clinics, Inc., buys $400,000 in medical supplies each year (at gross prices) from its major suppliers, Consolidated Services, which offers Langley terms of 2.5/10, net 45. Currently, Langley is paying the supplier...

Words: 1305 - Pages: 6

Premium Essay

Assignment 1 Hsa 525

...Assignment 1- Brandywine Homecare Brandywine Homecare, a not-for-profit business, had revenues of $12 million in 2007.  Expenses other than depreciation totaled 75 percent of revenues, which is $9 million, and depreciation expense was $1.5 million.  Income statement is the one of the three financial statements. The other two are the balance sheet and the statement of the cash flows. Brandywine Homecare’s total profit margin of 12.5 percent shows that the homecare makes 12.5 cents on every dollar of total revenues. If the company doubled its depreciation, both net income and profit margin would be zero. The cash flow will be the same for both cases, which is $3 million. Cash accounting and accrual accounting are two similar methods of maintaining accurate accounting records. Accrual accounting is considered to be the standard accounting practice for most companies. Equity is the ownership claim against total assets. For investor-owned businesses, equity is the amount of owner-supplied financing. Equity for not-for-profit businesses is the amount of capital supplied by the government grants, charitable contributions or other organization. Brandywine Homecare Brandywine Homecare, a not-for-profit business, has reported revenues of $12 million in 2007.  Expenses other than depreciation totaled 75 percent of revenues, and depreciation expense was $1.5 million.  It is reported that all revenues were collected in cash during the year and all expenses other than depreciation...

Words: 1469 - Pages: 6

Premium Essay

Hsa 525 Chapter 16, 17

...16.1 The checks that Park Place writes create disbursement float, which is “good” float. The clinic has $1,000 x 4 days = $4,000 in disbursement float. The checks that the clinic receives create collections float, which is “bad” float. The clinic has $1,000 x 3 days = $3,000 in collections float. The clinics net float is $1,000: Net float = Disbursement (good) float – Collections (bad) float = $4,000 – $3,000 = $1,000. In essence, the balance on the firm’s books is $1,000 less than the balance shown on the bank’s books. 16.2 a. The annual cost of operating the lockbox system is $103,350: Total cost = (Fixed cost per month x 12) + (Number of locations x Number of transfers x Cost per transfer) = ($6,500 x 12) + (10 x 260 x $9.75) = $78,000 + $25,350 = $103,350. b. The annual benefit is $97,500: Benefit = Number of days saved x Payments per day x Carrying cost = 3 x $325,000 x 0.10 = $97,500. c. In this situation, the annual cost of the lockbox system is greater than the annual benefit, and hence the lockbox system should not be initiated. 16.3 a. Seattle must pay the bills in 20 days if it wants to take advantage of the discount of three percent. If it wants to take the costly trade credit (forego the discount), it must pay the bills in 60 days. b. Using the approximate cost formula, the cost is 27.8%. Discount percent ...

Words: 1328 - Pages: 6

Premium Essay

Hsa 525 Week 9 Assignment 3

...Running Head: MERGER Merger Introduction There are certain organizations around the world which get engaged in the activities regarding merger. The health care organizations also get engaged in merger with other health care organizations. In the present paper, discussion shall be made regarding merger in health care organizations. There are certain financial drivers which are required to be considered by the health care organizations to move forward with merger. The financial planning and the success of the merger are to be taken into consideration by considering the financial statements of the organization after the merger has been carried out by the organizations. Financial Drivers Causing Merger There are three financial drivers which cause mergers within the health care organizations. The first financial driver causing merger is healthy corporate balance sheets. The second financial driver is low interest rates and inexpensive refinancing opportunities. The third financial driver is marginal growth of the organization (Buckley & Ghauri, 2002). All these financial drivers work in a positive manner for promoting mergers within the organization. In relation to merger, the organizations look at the potential health corporate balance sheets. In relation to the potential healthy balance sheets, the organizations look to get merged with another organization which has huge numbers in the balance sheet so that, the size of the balance...

Words: 1564 - Pages: 7

Premium Essay

Assignment # 4 – Medical Associates

...HSA 525-Health Financial Management Assignment # 4 – Medical Associates November 27, 2011 Medical Associates: Equity cost of capital, DCF, CAPM, risk, capital budgeting Medical Associates is a large for-profit group practice.  Its dividends are expected to grow at a constant rate of 7% per year into the foreseeable future.  The firm's last dividend (D0) was $2, and its current stock price is $23.  The firm's beta coefficient is 1.6; the rate of return on 20-year T-bonds currently is 9%; the expected rate of return is 13%.  The firm's target capital structure calls for 50% debt financing, the interest rate required on the business's new debt is 10%, and its tax rate is 40%.  1. Calculate Medical Associates' cost of equity estimate using the DCF method.  Using the DCF method  Cost of Equity = D1/P0 + g  D1 = expected dividend= D0 X (1+g) = 2 X 1.07 = 2.14  P0 current price = $23  g = growth rate = 7%  Cost of equity = 2.14/23 + 7% = 16.30%  2. Calculate the cost of equity estimate using CAPM.  Using CAPM  Cost of equity = Rf + (Rm-Rf) beta  Rf = risk free rate = 9%  Rm = return on market = 13%  beta = 1.6  Cost of equity = 9% + (13%-9%) 1.6 = 15.4%  3. On the basis of your answers to #1 & #2, what is your final estimate for the firm's cost of equity?  We take the average of the two costs as the final estimate  Final estimate is the average of the two = (16.3%+15.4%)/2 = 15.85%  4. Calculate the firm's estimate for corporate cost of capital.  Cost...

Words: 550 - Pages: 3

Premium Essay

Ruger Clinic

...ASSIGNMENT 2: HSA 525 Ruger Clinic Raymond Barber Dr. Japheth Kaluyu   1 What is the value of the cost pool? The Housekeeping department of Ruger Clinic consisted of $100,000 in total budgeted costs for the year of 2007. The cost pool is basically direct costs of one support department. In the case of Ruger Clinic, the cost pool's value is $100 K. 2aWhat is the allocation rate if patient services revenue is used as the cost driver? The cost pool, which must be found first, is the total cost of the Housekeeping department which is $100,000, while the patient services revenue was 5,000.00. The Allocation Rate equals the cost pool divided by the patient services revenue. That would be 100,000 divided by 5,000.00(100,000/5,000). The Allocation Rate is 0.02 for each of dollar of patient services revenue. = 100,000/ 5,000,000 = 0.02 for each dollar of patient services revenue. 2 b What is the allocation rate if hours of housekeeping services department is used as the cost driver? Allocation rate = cost pool/ hours of housekeeping department = 100,000/ 5000 = $20 per hour of services provided. 3 What is a cost-volume-profit (CVP) analysis and why is it useful to health services managers? Profit analysis is an analytical technique used to analyze the effects of volume changes on costs, and hence this analysis...

Words: 839 - Pages: 4