Free Essay

The Thermocouple

In:

Submitted By knasty
Words 553
Pages 3
Temporary adjustment of a theodolite

Similarly to the levelling instrument we can distinguish between temporary and permanent adjustment of the instrument. The temporary adjustments have to be carried out every time the theodolite is used.

2.1 Setting up, centring, and levelling up.

Setting up and centring concern the procedure necessary to place the instrument vertically above the station to be occupied (stationpoint). Levelling up is to bring the vertical axis V truly vertical. Very often theodolite surveying involves measurement of angles from a certain station point. The instrument should then be centred vertically above the stationmark and at the same time accurately levelled up. A suitable procedure to achieve this situation can be described in 7 points:

1. Put the tripod above the indicated stationpoint as accurately as possible, and with the top of the tripod approximately horizontal. The tripod legs are preferably at equal distances from the station mark. Now put the legs firmaly into the ground, so that that will be no danger of distrubance during measurement.
2. Mount the theodolite on top of the tripod, making sure that the instrument is firmly attached to it.
3. Use the optical plummet to determine if the mark of the stationpoint is within view. If not, repeat from point 1.
4. Use the three footscrews to get the stationmark exactly in probably not horizontal in this position.
5. Adjust one or more tripod legs to level the theodolite approximately. The circular bubble is used for this ‘approximate levelling up’.
6. Check if the thodolite is still correctly centered above the stationpoint. Small deviation can be adjusted by loosening the screw with which the theodolite is attached to the tripod and shifting the instrument. If the deviations are too large repeat from point 4.

fig. 6 Approximate levelling up procedure.

7. Carry out the ‘accurate levelling up’ procedure of the theodolite (including centring above a stationmark). The plate level is used for this.

fig. 7 Accurate levelling up procedure.

a) Rotate the alidate until the plate level is parallel to tow of the footscrews, position (a). Bring the bubble to the center of its run using these footscrews.
b) Rotate alidade over 90o clockwise to position (b) and use the third footscrew to center the bubble again.
c) Rotate again over 90o to position (c), and determine whether the bubble remains central. If not, the plate level is out of adjustment and half the deviation has to be taken away with the footscrews, both in position (a) and (b).
d) If the whole procedure has been carried out correctly, then the bubble should remain in the same position while rotating the alidade over 360o.
e) Check if the theodolite is still centered above your stationmark. If not, repeat from point 6.

2.2 Removing parallax.

Necessary to bring images of object and cross hairs in the same optical plane, making accurate readings possible.

Turn the focussing sleeve to the infinity mark.
Point to the sky or hold a white sheet of paper in front of the telescope for even illumination.
Turn the adjustment screw at eyepiece end until the cross hairs appear very sharp and distinct. Since the value of the dioptric scale depends upon the characteristics of the observer’s eye it can be set every time the observer uses this instrument.

Similar Documents

Premium Essay

Temperature and Response Time Measurement

...Temperature and Response time measurement Abstract In this experiment, a Resistive Thermal Device (RTD) and thermocouple were calibrated for first order measurement. Then, the response times of the RTD and thermocouple sensors were calculated. After performing this experiment, it was found that the time constants for the RTD were found to be -78.7 for cold to hot and 17.39 for hot to cold. The time constants for the thermocouple were found to be 3.519 for cold to hot and 2.64 for hot to cold. The rise times for the RTD were found to be 7.39 for cold to hot, and 7.9 for hot to cold. The rise times for the thermocouple were found to be 1.14 for cold to hot, and 1.27 for hot to cold. The error band for the RTD was found to be ±.0075, and the error band for the thermocouple was found to be ±.3887. Nomenclature T Time τ Time Constant E Error Band INTRODUCTION A Resistive Thermal Devices is a sensor used to measure the temperature by correlating the resistance of the RTD element with temperature. An RTD most commonly consists of a length of fine coiled wire wrapped around a glass or ceramic center. The element of the RTD is made of a pure material so its resistance can be calculated at various temperatures [1]. A thermocouple is that consists of two conductors which produce a voltage proportional to a temperature difference between either end of the pair of conductors. This allows for the measurement of the temperature at broad ranges...

Words: 1923 - Pages: 8

Premium Essay

Calibration of a Chromel-Alumel Thermocouple

...| Calibration of a Chromel-Alumel Thermocouple | | Module Title: Engineering ApplicationsModule Tutor: | Student Name: | Submission Date: | | Contents Objectives 3 Introduction 3 Apparatus 3 Theory 3 Results 4 Discussion 4 Conclusion 6 References 6 Calibration of a Chromel-Alumel Thermocouple Objectives * Use a thermometer to tell the changing temperatures of water * To become familiar with the principles of temperature measurement Introduction The purpose of this experiment was to measure the change in voltage in boiling water (electromotive force) against the temperature (measured by a thermometer) when cold water was added at stages. The voltage readings were taken using a Chromel-Alumel thermocouple. 1 6 5 2 Apparatus 1. Immersion heater and power supple 2. 500ml beaker half full with water 3. 3 1000ml beaker full with water 4. Thermometer 5. Thermocouple 6. Digital multimeter 7. Clamp and stand 7 4 Theory Refer to lab sheet for method used. A thermocouple is thermoelectric temperature sensor (which means it can measure temperature using electricity). A thermocouple consists of two wires that are made of different metals to one another which are coupled at the measurement junction and extended to the reference junction. [1] [1] A thermometer is “an instrument for measuring and indicating temperature, typically one consisting of a narrow, hermetically sealed glass tube marked...

Words: 957 - Pages: 4

Premium Essay

Nt1310 Unit 6 Lab Report

...Adedamola Iyiola M.E. LAB: Lab View Section 06   Lab View Adedamola Iyiola Department of Mechanical and Aerospace Engineering Rutgers University, Piscataway, New Jersey 08854 A Data acquisition board, thermocouple and a BNC terminal block were used simultaneously with Lab view to obtain measurements of different waveforms with varying frequencies of 500, 100, and 3000 Hz. Additionally, we varied sampling rates, at an input of 1000 Hz, at 500 Hz and 2500 Hz (0.5x and 2.5x the input wave frequency). Furthermore for the second set of measurements, we obtained data from running a thermocouple data acquisition program. We acquired temperature measurements of a room for a period of 60 seconds and an individual’s finger for 50 seconds....

Words: 1259 - Pages: 6

Premium Essay

Nt1310 Unit 5 Lab

...Before the lab began, a box with 2” styrofoam on the bottom and all sidewalls was created. A 2” styrofoam top with a small wooden part for additional cover was added. A lightbulb of unknown wattage was placed inside of the box and was turned on for an unknown time. A small hole was cut into one sidewall in order to place thermocouple probes into at a later time. The first portion of this lab was to set up the initial thermocouple setup, which was described in the apparatus section of this report. Project 7-A: After the apparatus was was set up, thermocouple probe tip voltage was recorded using NI DMM Soft Front Panel software. Voltage relating to ambient temperature was also recorded. Using the provided thermocouple tables and the law of intermediate temperatures, the temperature in the box was found....

Words: 430 - Pages: 2

Free Essay

Api 551

...Process Measurement Instrumentation API RECOMMENDED PRACTICE 551 FIRST EDITION, MAY 1993 American Petroleum Institute 1220 L Street, Northwest Washington, D.C. 20005 Process Measurement Instrumentation Manufacturing, Distribution and Marketing Department API RECOMMENDED PRACTICE 551 FIRST EDITION, MAY 1993 American Petroleum Institute SPECIAL NOTES 1. API PUBLICATIONS NECESSARILY ADDRESS PROBLEMS OF A GENERAL NATURE. WITH RESPECT TO PARTICULAR CIRCUMSTANCES, LOCAL, STATE, AND FEDERAL LAWS AND REGULATIONS SHOULD BE REVIEWED. 2. API IS NOT UNDERTAKING TO MEET THE DUTIES OF EMPLOYERS, MANUFACTURERS, OR SUPPLIERS TO WARN AND PROPERLY TRAIN AND EQUIP THEIR EMPLOYEES, AND OTHERS EXPOSED, CONCERNING HEALTH AND SAFETY RISKS AND PRECAUTIONS, NOR UNDERTAKING THEIR OBLIGATIONS UNDER LOCAL, STATE, OR FEDERAL LAWS. 3. INFORMATION CONCERNING SAFETY AND HEALTH RISKS AND PROPER PRECAUTIONS WITH RESPECT TO PARTICULAR MATERIALS AND CONDITIONS SHOULD BE OBTAINED FROM THE EMPLOYER, THE MANUFACTURER OR SUPPLIER OF THAT MATERIAL, OR THE MATERIAL SAFETY DATA SHEET. 4. NOTHING CONTAINED IN ANY API PUBLICATION IS TO BE CONSTRUED AS GRANTING ANY RIGHT, BY IMPLICATION OR OTHERWISE, FOR THE MANUFACTURE, SALE, OR USE OF ANY METHOD, APPARATUS, OR PRODUCT COVERED BY LETTERS PATENT. NEITHER SHOULD ANYTHING CONTAINED IN THE PUBLICATION BE CONSTRUED AS INSURING ANYONE AGAINST LIABILITY FOR INFRINGEMENT OF LETTERS PATENT. 5. GENERALLY, API STANDARDS ARE REVIEWED AND REVISED, REAFFIRMED, OR...

Words: 27566 - Pages: 111

Free Essay

Electrical Fundamentals

...Introduction Aircrafts require some form of electrical power to operate navigation, taxi, landing, strobe lights and other electronic systems. The electrical system consist of a battery and an alternator (you will find a generator on older type aircraft) to power the system and recharge the battery. You will also find fuses and switches and lights for indication purposes. Some electricity generators are used as backup power or other indication purposes. For example, a volt and/or ammeter are used for monitoring a change in voltage or current. However, this change in voltage or current could be brought about through the means of heat or light, which are electricity generators too. In a situation where primary power has failed, the aircraft storage batteries provide an emergency source of electrical power for operating electrical systems of an aircraft. All of these are connected through several meters (kilometres in large aircraft) of wire and connectors, attached to the airframe with insulation materials. In this report, I will be focusing on how these electricity generators are used in the various sections of the aircraft so as to understand the function they serve. I will be looking at four different methods of electricity generation. They are Magnetism and motion, Chemical Action, Light and Heat. Method of Electricity Generation 1 (Magnetism and Motion) It is important to note that magnetism itself does not create electricity. It requires...

Words: 1334 - Pages: 6

Premium Essay

Peltier Cell Lab Report

...device consist several of thermocouples. The thermocouple have two thermo elements that spreads between the hot side and cold side of the device. Fig. 1.1 • The efficiency of this device is around 5-8 % (Wikipedia.org).Generally, older devices are made from bimetallic junctions.it was very bulky. But, now a days highly doped semiconductors are used which is made from bismuth telluride (Bi2Te3) and lead telluride (PbTe). Practically Peltier cell working based on three effects 1) Peltier effect. 2) Thomson effect. 3) Seebeck effect. 1) Peltier Effect: • If we applied voltage to peltier unit then the heat is transferred from...

Words: 794 - Pages: 4

Premium Essay

Heat and Mass

...NATIONAL UNIVERSITY OF SCIENCE AND TECHNOLOGY PAKISTAN NAVY ENGINEERING COLLEGE HEAT & MASS TRANSFER LAB MANUAL ENGINEERING SCIENCES DEPARTMENT HEAT & MASS TRANSFER LAB MANUAL ENGINEERING SCIENCES DEPARTMENT LAB INSTRUCTIONS 1. No late submissions 2. If you miss the submission time, you will be marked zero in that lab 3. If cheating or plagiarism is observed in the lab report, zero will be marked in that lab. 4. Detailed calculations should be shown. 5. Neat work will earn good marks. 6. Graphs should be made by hands. MS excel graphs will not be accepted. LAB MARKS DISTRIBUTION LAB REPORT/ ATTENDENCE | 50% | LAB PERFORMANCE | 10% | MID TERM ASSESMENT/QUIZES | 10% | FINAL LAB ASSESMENT | 30 % | TOTAL | 100% | TABLE OF CONTENTS S.NO | DATE | OBJECTIVE | PAGE NO | SIGNATURE | 1. | | To show that the intensity of radiation on a surface is inversely proportional to the square of the distance of the surface from the source of radiation | | | 2. | | To show the intensity of radiation various as fourth power of source. | | | 3. | | Determination of barrier temperature gradient between two different metals in end to end pressure contact. | | | 4. | | To perform energy balance on air flowing throw a duct heated by cylindrical rod | | | 5. | | To perform energy balance on air flowing throw a duct heated by cylindrical rod using DAQ | | | 6. | | Determination...

Words: 4135 - Pages: 17

Premium Essay

Temperature Sensors

...Introduction Sensors- A sensor is a device that measures a physical quantity and converts it into a signal which can be read by an observer or by an instrument. Sensors are devices that are used to measure physical variables like temperature, pH, velocity, rotational rate, flow rate, pressure and many others. For example, a mercury thermometer converts the measured temperature into expansion and contraction of a liquid which can be read on a calibrated glass tube. A thermocouple converts temperature to an output voltage which can be read by a voltmeter. For accuracy, all sensors need to be calibrated against known standards. Today, most sensors do not indicate a reading on an analog scale (like a thermometer), but, rather, they produce a voltage or a digital signal that is indicative of the physical variable they measure. Those signals are often imported into computer programs, stored in files, plotted on computers and analyzed to death. Working The sensors work on the principle of analog to digital conversion. In order to collect data from the sensors, the signals from the sensors need to be converted into digital numbers that the processor can handle. The sensors generate a varying voltage based on what is measured and the processor cannot understand it. Remember, the processor is digital so it only understands ones and zeros. ● An analog¬to-digital converter or ADC allows a processor to measure voltages. The world outside the computer does not have discrete ...

Words: 1469 - Pages: 6

Premium Essay

Hrm 112 Class

...Prevention of Cross Contamination * Put a good personal hygiene program in place: To keep food handlers from contaminating food, operations need to have a good personal hygiene program. That program should include polices addressing critical hand practices like proper handwashing, hand care, and correct glove use. Staff cleanliness and work attire should also be addressed. This should focus on things like bathing, clean clothing, the proper use of hair restraints, and prohibited jewelry. Finally, policies should be put in place to make sure food handlers come to work healthy. These will address actions like reporting illnesses, covering wounds, and avoiding unsanitary habits. * Remind employees to wash their hands after touching anything that may have contaminated them. This is especially important after using the restroom and after handling raw meat, seafood, and poultry. After employees have washed their hands, make sure they don’t dry them on their aprons or any part of their uniforms. Instead they must use a single-use paper towel or hand dryer. Using a paper towel to avoid contact with faucet and door handles also will help minimize the spread of harmful viruses and bacteria. * Use separate equipment: Each type of food should be prepped and handled with a separate piece of equipment. For example, use one set of cutting boards, utensils, and containers for raw poultry. Use another set for raw meat, and use a third set for produce. Some operations use colored cutting...

Words: 970 - Pages: 4

Premium Essay

Effects of Study Habits on Academic Performance of ... Effects of Self-Study Habits on the Academic Achievement of Grade-Vi Pupils in English

...Electricity is produced in different ways from different sources. The most common sources are friction (static), heat (thermocouple), light (photo cell or solar cell), pressure (crystals), magnetism (generators) and chemical action (battery). 1. Static electricity is produced by friction. In friction, electrons in an insulator can be separated by rubbing to produce opposite charges that remain in the dielectric. 2. Heat energy is converted into electrical energy through the use of a device called thermocouple. It is a device consisting of two dissimilar metal wires twisted together at one end but open at the opposite end. When the joined end is heated, it produces a low DC voltage at the open end. The voltage produced by this device is directly proportional to the temperature of the metal wires. To produce a higher voltage, a number of thermocouples may be joined together. 3. Photoelectric matter. Some materials are photoelectric, which means that they can emit electrons when light strikes the surface. A device called photovoltaic cell or photo cell is made of a vacuum tube with two electrodes. One electrode is coated with sensitive materials such as cesium or selenium which give off electrons when illuminated. These electrons strike the other electrode which in turn, collects the electrons. If the tube is combined with a fixed voltage source, there will be electron flow in the circuit and this will activate photoelectric devices such as television cameral tubes, phototransistors...

Words: 397 - Pages: 2

Premium Essay

Convction

...THERMODYNAMICS Page 1 of 2 TD1005 Free and Forced Convection Shows free and forced convection from different heat transfer surfaces Works with Typical screenshot of the VDAS® software • Self-contained, bench-top mounting • Includes three of the most common heat transfer surfaces – flat plate, pinned and finned • Simple and safe to use • Thermocouples and a sensitive anemometer measure temperatures and air velocity – shown on a digital display • Additional hand-held thermocouple probe included – to measure temperatures along the length of the pins and fins of two heat transfer surfaces • Variable speed fan and variable power heat source for a range of tests • Can connect to TecQuipment’s Versatile Data Acquisition System (VDAS®) • • • • TecQuipment Ltd, Bonsall Street, Long Eaton, Nottingham NG10 2AN, UK T +44 115 972 2611 • F +44 115 973 1520 • E info@tecquipment.com • W www.tecquipment.com An ISO 9001 certified company VDAS is a registered trademark of TecQuipment Ltd DB/0411 Page 2 of 2 TD1005 Free and Forced Convection Description Experiments The bench-mounting equipment includes a vertical duct that holds the chosen heat transfer surface and all instruments needed. • Comparing free and forced convection for different surfaces TecQuipment include three different common heat transfer surfaces with the equipment: • A Flat Plate • Comparison of free convection from vertical and horizontal (finned) surfaces ...

Words: 744 - Pages: 3

Free Essay

Deposition Technique

...California Institute of Technology Physics 77 Vacuum Techniques and Thin Film Deposition Experiment 3 (October 2001) 1 Introduction Much of modern experimental physics is done under vacuum. Design and construction of vacuum apparatus is one of the most useful ”bread and butter” skills an experimentalist in condensed matter, atomic, or optical physics can have, and the subject of vacuum engineering is a vast one. This lab serves as an introduction to basic vacuum techniques and thin film growth, another often essential skill for condensed matter physicists. This lab is an optional prerequisite for Experiment 10, Condensed Matter Physics at Cryogenic Temperatures, for which you can grow your own samples for Weak Localization measurements if you choose. 2 Pressure and gas flow In vacuum work, pressures are almost always measured in millimeters of mercury, or torr. One torr is just the pressure necessary to support a column of mercury with a height of one millimeter. The conversion to units more familiar to readers of physics textbooks is 1atmosphere = 101kPa = 760torr There are two pressure regimes of interest to the scientist working with vacuum systems, and gases behave differently in each regime. The first, the viscous flow regime, describes the case where gas flows as a fluid, where the mean free path of the gas molecules is much smaller than the dimensions of the apparatus. The second, the molecular flow regime, describes the high-vacuum case, where the mean...

Words: 3525 - Pages: 15

Premium Essay

Comp

...lts and Discussion The heart of a laboratory report is the presentation of the results and the discussion of those results. In some formats, "Results" and "Discussion" appear as separate sections. However, P.B. Medawar [1979] makes a strong case that the two should appear together, particularly when you have many results to present (otherwise, the audience is faced with a "dump" of information that is impossible to synthesize). Much here depends upon your experiment and the purpose of your laboratory report. Therefore, pay attention to what your laboratory instructor requests. Also, use your judgment. For instance, combine these sections when the discussion of your first result is needed to understand your second result, but separate these sections when it is useful to discuss the results as a whole after all results are reported. In discussing the results, you should not only analyze the results, but also discuss the implications of those results. Moreover, pay attention to the errors that existed in the experiment, both where they originated and what their significance is for interpreting the reliability of conclusions. One important way to present numerical results is to show them in graphs. Sample Results and Discussion This section analyses the results of the experiment. The experiment went as expected with no unusual events that would have introduced error. The voltages as measured for the pressure and temperature transducers appear in Table A-1 of the Appendix...

Words: 562 - Pages: 3

Free Essay

Term Paper Me3281

...ME3281 Microsystems Design and Applications TERM PAPER DEPARTMENT OF MECHANICAL ENGINEERING MEMS Energy Harvesters LIM HUI HUA ALVINA A0083044M 2014 Table of Contents 1. Introduction: 2 2. Brief History of Electricity Transduction 2 3. Types of Micro Energy Harvesters 3 3.1 Energy Harvesting from Vibration 3 3.1.1 Fabrication Techniques 5 3.1.2 Applications, Challenges and the Future 6 3.2 Energy Harvesting from Thermal Sources 6 3.2.1 Fabrication Techniques 8 3.2.2 Applications, Challenges and the Future. 9 3.3 Energy Harvesting from Electromagnetic Waves 10 3.3.1 Applications, Challenges and the Future 11 3.4 Energy Harvesting from Light Sources 11 3.4.1 Fabrication 12 3.4.2 Applications, Challenges and the Future 13 4. Conclusion 13 References: 14 1. Introduction: One of the goals of engineers and scientists in this already tech-savvy age is to be able to design a device that is capable of powering itself for its lifetime without having to replace or recharge its battery using a power chord. These allow remote devices to be placed in hostile or inaccessible environments without requiring any or little maintenance such as the changing of batteries. This is especially applicable for silicon-based electronics, such as biomedical implants that have low power consumption, where batteries will largely affect its size; operational cost of the device, or perhaps even release harmful chemicals into the body. In addition, wireless sensor...

Words: 4736 - Pages: 19