Free Essay

Cars

In:

Submitted By tojmage
Words 12102
Pages 49
USER Manual

1038585 JR 10/02/06

Important!

Fill in the information below when you receive the BiPAP Auto M Series system.
Serial No.: _______________________________ (located on the bottom of the device) System Prescribed for: __________________________________________ Date of Purchase or Rental: ______________________________________ Pressure Setting: _____ cm H2O Mask Type: __________________________________________________ Mask Size: ___________________________________________________

If you have any questions concerning the system, contact: • Home Care Company: _______________________________________ • Telephone Number: _________________________________________ Telephone Number: _________________________________________ Respironics, Inc. 1001 Murry Ridge Lane Murrysville, Pennsylvania 15668-8550 USA Customer Service Telephone Number: 1-800-345-6443 (US and Canada) or 1-724-387-4000 • Health Care Professional: _____________________________________

The BiPAP® Auto M Series with Bi-Flex® system is covered by one or more of the following patents: 5,148,802; 5,313,937; 5,433,193; 5,632,269; 5,803,065; 6,029,664; 6,305,374; 6,539,940; 5,535,738; 5,794,615; 6,105,575; 6,609,517; 6,837,260; 5,645,035; 6,286,508; 6,550,478; 6,752,150; 6,615,831 5,970,975. Other patents pending. BiPAP, Whisper Swivel, Encore Pro, and Encore Pro SmartCard are trademarks of Respironics, Inc. © 2006 Respironics, Inc. and its affiliates. All rights reserved.

Complete the following steps to set up your M Series device.

Plug the AC power cord connector into the power supply, then plug the power cord into an electrical outlet.

Plug the power supply cord connector into the device.

Attach the tubing to the optional swivel and attach the swivel to the device’s air outlet. Or, attach the tubing directly to the air outlet.

Press the therapy.

button to begin

Complete the following steps to set up your M Series device and humidifier.

Connect the humidifier power supply connector to the device power supply. Plug the AC power cord connector into the power supply, then plug the power cord into an electrical outlet.

Plug the power supply cord connector into the device.

Attach the tubing to the optional swivel and attach the swivel to the device’s air outlet. Or, attach the tubing directly to the air outlet.

Lift the humidifier door to access the water tank.

Fill the tank with water until it reaches the fill line.

Place the water tank back in the humidifier.

CAUTION
DO NOT TIP OR TRANSPORT THE DEVICE WITH WATER IN THE HUMIDIFIER TANK.

Press the therapy.

button to begin

Turn the humidifier knob to the desired setting.

Final system setup for the device.

Final system setup for the device and humidifier.

taBle of ContentS
Chapter 1: Introduction ..................................................................................................................1-1 1.1 System Contents ........................................................................................................1-1 1.2 Intended Use ...............................................................................................................1-2 1.3 Warnings, Cautions, and Contraindications .....................................................1-2 1.3.1 Warnings ...........................................................................................................1-2 1.3.2 Cautions ............................................................................................................1-4 1.3.3 Contraindications ..........................................................................................1-4 1.4 System Overview .......................................................................................................1-5 1.4.1 Breathing Circuit Overview........................................................................1-6 1.5 Glossary .........................................................................................................................1-7 1.6 Symbol Key ..................................................................................................................1-8 1.7 How to Contact Respironics ..................................................................................1-8 Chapter 2: Device Controls and Displays .................................................................................2-1 2.1 Controls and Displays ..............................................................................................2-1 2.1.1 Control Panel Inactivity ...............................................................................2-2 2.2 Rear Panel .....................................................................................................................2-2 Chapter 3: Setup................................................................................................................................3-1 3.1 Installing the Air Filters ............................................................................................3-1 3.2 Where to Place the Device......................................................................................3-2 3.3 Connecting the Breathing Circuit .......................................................................3-2 3.4 Supplying Power to the Device ............................................................................3-4 3.4.1 Using AC Power .............................................................................................3-4 3.4.2 Using DC Power .............................................................................................3-5 3.5 Complete Assembly Example ...............................................................................3-6 Chapter 4: Device Operation ........................................................................................................4-1 4.1 Available Therapies ...................................................................................................4-1 4.2 Starting the Device ...................................................................................................4-1 4.3 Using the Ramp and Bi-Flex Features.................................................................4-3 4.3.1 Ramp Feature..................................................................................................4-3 4.3.2 Bi-Flex Comfort Feature ..............................................................................4-3 4.3.2.1 Bi-Flex Enabled ...................................................................................4-3 4.3.2.2 Bi-Flex Locked.....................................................................................4-4 4.3.2.3 Bi-Flex Disabled .................................................................................4-4 BIpap auto m SerIeS uSer manual i

4.4 Changing the Device Settings .............................................................................. 4-4 4.4.1 Navigating the Display Screens................................................................ 4-5 4.4.1.1 Viewing Data on the Patient Data Screens............................... 4-5 4.4.1.2 Viewing and Modifying Patient Setup Screens ...................... 4-7 4.5 Completing the FOSQ Questionnaire ................................................................ 4-9 4.6 Patient Reminder Screen ......................................................................................4-10 Chapter 5: Device Alerts and Troubleshooting ...................................................................... 5-1 5.1 Device Alerts ............................................................................................................... 5-1 5.2 Troubleshooting......................................................................................................... 5-3 Chapter 6: Accessories .................................................................................................................... 6-1 6.1 Adding a Humidifier ................................................................................................. 6-1 6.2 Using the SmartCard .............................................................................................. 6-2 6.3 Adding Supplemental Oxygen ............................................................................. 6-2 Chapter 7: Cleaning and Maintenance...................................................................................... 7-1 7.1 Cleaning the Device ................................................................................................. 7-1 7.2 Cleaning or Replacing the Filters ......................................................................... 7-1 7.3 Cleaning the Tubing ................................................................................................. 7-3 7.4 Service ........................................................................................................................... 7-3 7.5 Traveling with the System ...................................................................................... 7-3 7.5.1 International Travel ....................................................................................... 7-4 Chapter 8: Specifications................................................................................................................ 8-1 Appendix A: EMC Information......................................................................................................A-1

ii

BIpap auto m SerIeS uSer manual

Chapter 1: IntroduCtIon
This chapter provides information on: • • • • • • System contents Intended Use Warnings, cautions, and contraindications System overview Glossary and symbol key How to contact Respironics

1.1 SyStem ContentS
Your BiPAP® Auto M Series system includes the following items:
User Manual Carrying Case

Flexible Tubing

Power Supply

Reusable Gray Foam Filter Disposable Ultra-fine Filter
B i-F le x

AC Power Cord BiPAP Auto M Series Device

fIgure 1–1 SyStem ContentS Note: If your system includes a humidifier, you will receive additional items with your package. See the instructions included with your humidifier for more information. If any of the above items are missing, contact your home care provider.

Note:

BIpap auto m SerIeS uSer manual

1-1

1.2 Intended uSe
The Respironics BiPAP® Auto with Bi-Flex® M Series system delivers positive airway pressure therapy for the treatment of Obstructive Sleep Apnea (OSA) only in spontaneously breathing patients weighing >66 lbs (30 kg). The device is to be used only on the instruction of a licensed physician. The system can deliver Bi-level therapy or Auto Bi-level therapy, both with and without Bi-Flex. Your home care provider will make the correct pressure settings according to your health care professional’s prescription. When set in the Auto Bi-level therapy mode, the system will monitor your breathing as you sleep and automatically adjust the pressure to meet your needs. When in Bi-level therapy, the system will deliver a continuous, set pressure during the night. Several accessories are available to make your OSA treatment with the BiPAP Auto with Bi-Flex M Series system as convenient and comfortable as possible. To ensure that you receive the safe, effective therapy prescribed for you, use only Respironics accessories.

1.3 WarnIngS, CautIonS, and ContraIndICatIonS
Caution: U.S. federal law restricts this device to sale by or on the order of a physician.

1.3.1 WarnIngS
A warning indicates the possibility of injury to the user or the operator. • • • This manual serves as a reference. The instructions in this manual are not intended to supersede the health care professional’s instructions regarding the use of the device. The operator should read and understand this entire manual before using the device. The device should be used only with masks and connectors recommended by Respironics or with those recommended by the health care professional or respiratory therapist. A mask should not be used unless the device is turned on and operating properly. The exhalation port(s) associated with the mask should never be blocked. Explanation of the Warning: The device is intended to be used with special masks or connectors that have exhalation ports to allow continuous flow of air out of the mask. When the device is turned on and functioning properly, new air from the device flushes the exhaled air out through the mask exhalation port. However, when the device is not operating, enough fresh air will not be provided through the mask, and exhaled air may be rebreathed. Rebreathing of exhaled air for longer than several minutes can in some circumstances lead to suffocation. Use only the breathing circuit provided by your home care provider. When using a breathing circuit that contains a mask with an integrated exhalation port or a circuit with a separate exhalation device, do not tape, seal, or otherwise block the vent openings. Doing so could result in suffocation. BIpap auto m SerIeS uSer manual

• •

1-2



If oxygen is used with the device, the oxygen flow must be turned off when the device is not in use. Explanation of the Warning: When the device is not in operation and the oxygen flow is left on, oxygen delivered into the tubing may accumulate within the device’s enclosure. Oxygen accumulated in the device enclosure will create a risk of fire. Oxygen supports combustion. Oxygen should not be used while smoking or in the presence of an open flame. When using oxygen with this system, a Respironics Pressure Valve (Part Number 302418) must be placed in-line with the patient circuit. Failure to use the pressure valve could result in a fire hazard. Do not use the device in the presence of a flammable anaesthetic mixture in combination with oxygen or air, or in the presence of nitrous oxide. Do not use this device if the room temperature is warmer than 95°F (35°C). If the device is used at room temperatures warmer than 95°F (35°C), the temperature of the airflow may exceed 106°F (41°C). This could cause irritation or injury to your airway. Do not operate the device in direct sunlight or near a heating appliance because these conditions can increase the temperature of the air coming out of the device. Do not attempt to wear your mask without the device turned on. Doing so could result in CO2 rebreathing. Contact your health care professional if symptoms of sleep apnea recur. If you notice any unexplained changes in the performance of this device, if it is making unusual or harsh sounds, if the device or the power supply are dropped or mishandled, if water is spilled into the enclosure, or if the enclosure is broken, discontinue use and contact your home care provider. Repairs and adjustments must be performed by Respironics-authorized service personnel only. Unauthorized service could cause injury, invalidate the warranty, or result in costly damage. Periodically inspect electrical cords, cables, and the power supply for damage or signs of wear. Discontinue use and replace if damaged. To avoid electric shock, unplug the device before cleaning it. DO NOT immerse the device in any fluids. Pins of connectors identified with the ESD warning symbol ( )should not be touched. Connections should not be made to these connectors unless ESD precautionary procedures are used. Precautionary procedures include methods to prevent build-up of electrostatic discharge (e.g., air conditioning, humidification, conductive floor coverings, non-synthetic clothing), discharging one’s body to the frame of the equipment or system or to earth or a large metal object, and bonding oneself by means of a wrist strap to the equipment or system or to earth.

• •

• •

• • • •



• • •

BIpap auto m SerIeS uSer manual

1-3

1.3.2 CautIonS
A Caution indicates the possibility of damage to the device. • • • • • The device may only be operated at temperatures between 41°F (5°C) and 95°F (35°C). If this device has been exposed to either very hot or very cold temperatures, allow it to adjust to room temperature before starting therapy. Condensation may damage the device. Do not immerse the device or allow any liquid to enter the enclosure or the inlet filter. A properly installed, undamaged reusable foam inlet filter is required for proper operation. Tobacco smoke may cause tar build-up within the device, which may result in the device malfunctioning. Additional warnings, cautions, and notes are located throughout this manual.

Note:

1.3.3 ContraIndICatIonS
When assessing the relative risks and benefits of using this equipment, the clinician should understand that this device can deliver pressures up to 25 cm H2O. In the event of certain fault conditions, a maximum pressure of 35 cm H2O is possible. Studies have shown that the following pre-existing conditions may contraindicate the use of CPAP therapy for some patients: • • • • • Bullous Lung Disease Pathologically Low Blood Pressure Bypassed Upper Airway Pneumothorax Pneumocephalus has been reported in a patient using nasal Continuous Positive Airway Pressure. Caution should be used when prescribing CPAP for susceptible patients such as those with: cerebral spinal fluid (CSF) leaks, abnormalities of the cribriform plate, prior history of head trauma, and/or pneumocephalus. (Chest 1989; 96:1425-1426)

The use of positive airway pressure therapy may be temporarily contraindicated if you exhibit signs of a sinus or middle ear infection. Not for use with patients whose upper airways are bypassed. Contact your physician if you have any questions concerning your therapy.

1-4

BIpap auto m SerIeS uSer manual

1.4 SyStem overvIeW
The BiPAP Auto M Series system offers several options in how therapy is delivered, so treatment can be personalized to meet your needs. The system delivers two different positive pressure levels: IPAP (Inspiratory Positive Airway Pressure) and EPAP (Expiratory Positive Airway Pressure). Your home care provider will make the correct pressure settings. When prescribed for you, the device provides several special features to help make your therapy more comfortable. The ramp function allows you to lower the pressure when you are trying to fall asleep. The air pressure will gradually increase until your prescription pressure is reached. You also have the option of not using the ramp feature at all. Additionally, the Bi-Flex comfort feature provides you with pressure relief when you exhale during therapy. Several accessories are also available for use with your device. Contact your home care provider to purchase any accessories not included with your system. The device, shown in Figure 1–2, supplies air pressure through a breathing circuit.

fIgure 1–2 BIpap auto m SerIeS devICe Figure 1–3 illustrates many of the device features, described in the following table.
Ramp Button Start/Stop Button Bi-Flex Button Display Screen (under door) le x

Power Inlet

Air Outlet

Medical Equipment Note (on bottom)

Accessory Module

Filter Area

fIgure 1–3 SyStem overvIeW BIpap auto m SerIeS uSer manual 1-5

devICe feature Accessory Module Air Outlet Bi-Flex Button

deSCrIptIon If applicable, insert the optional accessory card here. Connect the flexible tubing here. This button allows you to change the Bi-Flex comfort setting, if Bi-Flex is prescribed for you. See Chapter 4 for more information about Bi-Flex. Shows therapy settings, patient data, instructions, and error messages. A reusable, gray foam filter must be placed in the filter area to screen out normal household dust and pollens. An optional, white ultra-fine filter can also be used for more complete filtration of very fine particles. For ease at airport security stations, there is a note on the bottom of the device stating that it is medical equipment. It may help if you also take this manual with you when you travel. Connect the power cord here. This button starts or restarts the ramp cycle. This button starts or stops the airflow.

Display Screen Filter Area

Medical Equipment Note

Power Inlet Ramp Button Start/Stop Button

1.4.1 BreathIng CIrCuIt overvIeW
The patient breathing circuit, shown in Figure 1–4, consists of the following: • Circuit tubing to deliver air from the device to your interface (e.g., mask) • A mask or other patient interface device to deliver the prescribed pressure to your nose or nose and mouth, depending on which interface has been prescribed for you • An exhalation device to vent exhaled air from the circuit
Patient Interface

Exhalation Device Circuit Tubing

Exhalation Port

Mask's Connector Flexible Tubing Connector

Circuit with Separate Exhalation Device

fIgure 1–4 typICal BreathIng CIrCuItS

Circuit with Mask with Integrated Exhalation Port

Note:

The exhalation port may be part of the interface or may be part of a separate exhalation device, but is required to minimize the potential for CO2 rebreathing. BIpap auto m SerIeS uSer manual

1-6

1.5 gloSSary
The following terms and acronyms appear throughout this manual: term/aCronym Active State Apnea Auto Bi-level Auto-Off Auto-On BPM Bi-Flex Compliance Check Value EPAP FOSQ defInItIon The state of the device when power is applied, the airflow is on, and the device is providing therapy. A condition marked by the cessation of spontaneous breathing. Adjusts the pressure automatically to improve patient comfort based on monitoring of apnea, hypopnea, and snoring events. When this feature is enabled, the device automatically discontinues therapy whenever the mask is removed. With this feature, the device automatically initiates therapy when you begin breathing on the device. This feature is always enabled. Breaths Per Minute A therapy feature that establishes a level of pressure relief taking place at the end of inhalation and at the start of exhalation. Represents a code used by the home care provider to determine the accuracy of reported data. Expiratory Positive Airway Pressure Functional Outcomes of Sleep Questionnaire – A “quality of life” questionnaire designed specifically for people with sleep disorders. The results allow health care professionals to assess how CPAP therapy has improved the quality of their patients’ lives. Inspiratory Positive Airway Pressure Liters Per Minute Obstructive Sleep Apnea The display mode in which you can view certain stored information, such as therapy usage hours. The display mode in which you can change patient-adjustable device settings, such as the ramp starting pressure. A feature that may increase patient comfort when therapy is started. The ramp feature reduces pressure and then gradually increases the pressure to the prescription setting so you can fall asleep more comfortably. The state in which the device does not provide therapy. The device enters this state if an error is detected. The state of the device when power is applied but the airflow is turned off.

IPAP LPM OSA Patient Data Menu Patient Setup Menu Ramp

Safe State Standby State

BIpap auto m SerIeS uSer manual

1-7

1.6 SymBol Key
The following symbols appear on the device and power supply: SymBol defInItIon Consult accompanying instructions for use. DC Power Type BF Applied Part Class II (Double Insulated)

IPX1

Drip Proof Equipment Electrostatic Discharge European Declaration of Conformity Canadian/US Certification Notified Body Approval for Standards Compliance TUV Safety Standard Compliance UL Recognized for Canada and the United States No User Serviceable Parts

1.7 hoW to ContaCt reSpIronICS
To have your device serviced, contact your home care provider. If you need to contact Respironics directly, call the Respironics Customer Service department at 1-800-345-6443 or 1-724-387-4000. You can also use the following address:

Visit Respironics web site at: www.respironics.com 1-8 BIpap auto m SerIeS uSer manual

Chapter 2: devICe ControlS and dISplayS
This chapter describes the device’s control buttons and displays, patient circuit connections, and rear panel connections.

2.1 ControlS and dISplayS
Figure 2–1 shows the three primary control buttons on the BiPAP Auto M Series device.
Bi-Flex

fIgure 2–1 prImary Control ButtonS These buttons are described below. Button deSCrIptIon Ramp – When the airflow is on, this button allows you to activate or restart the ramp function. Ramp lowers the airflow pressure and then gradually increases it, allowing you to fall asleep more easily. Start/Stop – This button starts the device’s airflow and places the device in the Active state, or stops the airflow, and places the device in Standby. You can also press this button to exit any setting screen. Bi-Flex Bi-Flex – This button adjusts the Bi-Flex comfort setting. If you do not want to modify the settings that your home care provider has set for you, you only need these three buttons to start and stop therapy and do not need to use the display buttons located under the device cover door. Figure 2–2 shows the device display and navigation buttons that are under the cover door.

– FOSQ

Setup +

fIgure 2–2 dISplay ButtonS The display buttons are described below: Button deSCrIptIon Allows you to access the Data screens. Also allows you to navigate to the previous screen when in the Data or Setup menu. Allows you to access the Data screens. Also allows you to navigate to the next screen when in the Data or Setup menu. Allows you to enter the FOSQ screens. Also allows you to decrease the settings on the setup screens. Allows you to enter the Setup screens. Also allows you to increase the settings on the setup screens. BIpap auto m SerIeS uSer manual 2-1

The device’s display screen shows the set pressure, patient data, instructions, and error messages. See Chapter 4 for instructions on navigating the display screens. Figure 2–3 provides a view of the entire device control panel, with the cover door open.
Placement Area for Optional Quick Reference Label Cover Door Previous Screen Button Decrease Button
Bi-Flex

Next Screen Button Increase Button

Ramp Button

Start/Stop Button

Bi-Flex Button

fIgure 2–3 devICe Control panel

2.1.1 Control panel InaCtIvIty
Some screens have time-out periods. The screen’s timer starts when the screen is initially displayed and is restarted whenever a button is pressed. The screen times out after one minute if there is no activity and returns to the Standby screen.

2.2 rear panel
Figure 2–4 shows the device’s rear panel (without a humidifier).
DC Power Inlet

Air Outlet Port

Accessory Slot

fIgure 2–4 rear panel The rear panel contains the following: • An accessory slot for optional accessories such as the SmartCard or modem (see Chapter 6, Accessories, for more information). • A filter area where the filter(s) supplied with your device should be inserted. • A DC power inlet where the power cord is connected (see Chapter 3, Setup, for complete information on supplying power to the device). • The air outlet port where the breathing circuit’s flexible tubing is attached. 2-2 BIpap auto m SerIeS uSer manual

Filter Area

Chapter 3: Setup
This chapter provides instructions on how to: • Install the air filters • Position the device • Connect the breathing circuit • Supply power to the device

3.1 InStallIng the aIr fIlterS
Caution: A properly installed, undamaged foam filter is required for proper operation. The device uses a gray foam filter that is washable and reusable, and an optional white ultra-fine filter that is disposable. The reusable filter screens out normal household dust and pollens, while the optional ultra-fine filter provides more complete filtration of very fine particles. The gray reusable filter must be in place at all times when the device is operating. The white ultra-fine filter is recommended for people who are sensitive to tobacco smoke or other small particles. Two reusable gray foam filters and one disposable white ultra-fine filter are supplied with the device. If your filters are not already installed when you receive your device, you must at least install the reusable gray foam filter before using the device. To install the filter(s): 1. If you are using the white disposable ultra-fine filter, insert it into the filter area first, meshside facing in, towards the device. 2. Insert the gray foam filter into the filter area as shown in Figure 3–1. Note: If you are not using the white disposable filter, simply insert the gray foam filter into the filter area.
Filter Area

Disposable Ultra-Fine Filter (optional)

Reusable Gray Foam Filter (required)

fIgure 3–1 InStallIng the aIr fIlter Note: See Chapter 7, Cleaning and Maintenance, for information on how to clean or replace the air filters. 3-1

BIpap auto m SerIeS uSer manual

3.2 Where to plaCe the devICe
Place the device on a firm, flat surface somewhere within easy reach of where you will use it. Make sure the filter area on the back of the device is not blocked by bedding, curtains, or other items. Air must flow freely around the device for the system to work properly. Make sure the device is away from any heating or cooling equipment (e.g., forced air vents, radiators, air conditioners).

3.3 ConneCtIng the BreathIng CIrCuIt
To use the system, you will need the following accessories in order to assemble the recommended circuit: • • • Respironics interface (e.g, nasal mask) with integrated exhalation port (or Respironics interface with a separate exhalation device such as the Whisper Swivel® II) Respironics 6 ft. (1.83 m) flexible tubing (with optional swivel) Respironics headgear (for the patient interface) If the device is used by multiple persons (e.g., rental devices), a lowresistance, main flow bacteria filter should be installed in-line between the device and the circuit tubing to prevent contamination.

Warning:

To connect your breathing circuit to the device, complete the following steps: 1. Connect the flexible tubing to the air outlet on the back of the device, as shown in Figure 3–2. You can use the optional, detachable swivel that is provided with your device and already attached to the tubing (shown in Figure 3–2), or you can remove the swivel and connect the tubing directly to the air outlet.

Detachable Swivel

fIgure 3–2 ConneCtIng the flexIBle tuBIng Note: If required, connect a bacteria filter to the device air outlet, and then connect the flexible tubing to the outlet of the bacteria filter.

3-2

BIpap auto m SerIeS uSer manual

2. Connect the tubing to the mask: a. If you are using a mask with a built-in exhalation port, connect the mask’s connector to the flexible tubing, as shown in Figure 3–3.

Exhalation Port Mask's Connector Flexible Tubing Connector

fIgure 3–3 ConneCtIng a maSK WIth BuIlt-In exhalatIon port b. If you are using a mask with a separate exhalation device, connect the open end of the flexible tubing to the exhalation device as shown in Figure 3–4. Position the exhalation device so that the vented air is blowing away from your face. Connect the mask’s connector to the exhalation device.

Exhalation Device

Flexible Tubing Connector

fIgure 3–4 ConneCtIng a maSK WIth a Separate exhalatIon devICe Warning: The exhalation device (e.g., Whisper Swivel II) or exhalation port (on masks with an integrated exhalation port) is designed to exhaust CO2 from the patient circuit. Do not block or seal the ports on the exhalation device. If you are using a full face mask (i.e., a mask covering both your mouth and your nose), the mask must be equipped with a safety (entrainment) valve.

Warning:

3. Attach the headgear to the mask. See the instructions that came with your headgear. BIpap auto m SerIeS uSer manual 3-3

3.4 SupplyIng poWer to the devICe
You can power the device using AC or DC power. Caution: If this device has been exposed to either very hot or very cold temperatures, allow it to adjust to room temperature before beginning the following setup procedures. Route the wires to avoid tripping. This device is activated when the power cord is connected. Pressing the button turns the airflow on or off. If you are using your device with a humidifier, refer to the Quickstart Guide included with your device or the instructions included with your humidifier for details on how to power the device and humidifier.

Warning: Warning: Important!

3.4.1 uSIng aC poWer
Complete the following steps to operate the device using AC power. 1. Plug the socket end of the AC power cord into the power supply, as shown in Figure 3–5.

fIgure 3–5 ConneCtIng the poWer Cord to the poWer Supply 2. Plug the pronged end of the AC power cord into an electrical outlet that is not controlled by a wall switch. 3. Plug the power supply cord’s connector into the power inlet on the back of the device, as shown in Figure 3–6.

3-4

BIpap auto m SerIeS uSer manual

fIgure 3–6 ConneCtIng the poWer Supply Cord to the devICe 4. Ensure that all connections are secure. Important! Warning: To remove AC power, disconnect the power supply cord from the electrical outlet. Inspect the power cord often for any signs of damage. Replace a damaged power cord immediately.

3.4.2 uSIng dC poWer
The Respironics DC Power Cord or Portable Battery Pack can be used to operate this device in a stationary recreational vehicle, boat, or motor home. The Respironics DC Battery Adapter Cable (Reorder Number 532209), when used with the DC Power Cord, enables the device to be operated from a 12 VDC free-standing battery. Caution: When DC power is obtained from a vehicle battery, the device should not be used while the vehicle’s engine is running. Damage to the vehicle or the device may occur. Caution: Only use a Respironics DC Power Cord, Portable Battery Pack, and Battery Adapter Cable. Use of any other system may cause damage to the device or vehicle. Refer to the instructions supplied with the DC Power Cord, Portable Battery Pack and Battery Adapter Cable for information on how to operate the device using DC power.

BIpap auto m SerIeS uSer manual

3-5

3.5 Complete aSSemBly example
Figure 3–7 shows an example of how a complete assembly will look, with breathing circuit connected and power applied to the device.

fIgure 3–7 fInal aSSemBly example (ShoWn WIth optIonal aCCeSSory module) Figure 3–8 shows an example of how you should route your tubing and situate your device on your nightstand for the best setup possible. This will help prevent the device from falling off your nightstand or table.

fIgure 3–8 reCommended devICe and tuBIng plaCement

3-6

BIpap auto m SerIeS uSer manual

Chapter 4: devICe operatIon
This chapter explains the therapies available with the BiPAP Auto M Series device, how to start the device, and how to change the settings.

4.1 avaIlaBle therapIeS
The device delivers the following therapies: • Bi-level – Provides one level of output pressure during EPAP (Expiratory Positive Airway Pressure) and a second higher level during IPAP (Inspiratory Positive Airway Pressure). If your health care professional prescribed ramp for you, you can press the Ramp button to reduce the pressure and then gradually increase the pressure to the therapeutic pressure setting so that you can fall asleep more comfortably. • Bi-level with Bi-Flex – Bi-level therapy with pressure relief upon exhalation to improve patient comfort based on patient needs. • Auto Bi-level – Delivers spontaneous Bi-level therapy while automatically adjusting EPAP and IPAP levels to meet the patient’s needs. • Auto Bi-level with Bi-Flex – Auto Bi-level therapy with pressure relief upon exhalation to improve patient comfort based on patient needs.

4.2 StartIng the devICe
1. Plug the device in to an AC or DC power source. The three primary buttons light up and the Software Version screen momentarily appears, shown in Figure 4–1.

fIgure 4–1 SoftWare verSIon SCreen Note: Version 1.0 shown in Figure 4–1 is an example. Your device may have a different software version installed.

V 1.0

2. The next screen to appear is the Standby (or Inactive) screen, shown in Figure 4–2. You can access the FOSQ questionnaire or patient settings and data from this screen. See Section 4.3 for instructions on how to navigate the display screens.

fIgure 4–2 StandBy SCreen Note: The FOSQ option will only appear on the Standby screen if a SmartCard is inserted in the device or if a modem is connected. 4-1

– FOSQ

Setup +

BIpap auto m SerIeS uSer manual

3. Press the button to turn on the airflow. The Active Display screen appears, shown in Figure 4–3.

9.0

E

cmH2O

20.0

I

cmH2O

Active Screen During Exhalation (EPAP)

Active Screen During Inhalation (IPAP)

fIgure 4–3 aCtIve dISplay (WIth SmartCard) The Active Display screen shows the set pressure. Some symbols indicating when certain conditions are turned on or detected will also appear on the Active Display screen or on the Standby screen. These symbols are described below: SymBol deSCrIptIon Ramp – You can initiate the ramp feature by pressing the Ramp button. The ramp symbol displays on the Active Display screen when the ramp function is active. E I Mask Leak Auto Off EPAP – Expiratory Positive Airway Pressure IPAP – Inspiratory Positive Airway Pressure Mask Leak Alert – This flashing text displays on the Active Display screen if the Mask Alert setting is enabled and the device detects an excessive mask leak. Auto-Off – This flashing text displays on the Standby screen if the Auto-Off setting is enabled and the device detects a mask disconnect. SmartCard – This symbol displays if a SmartCard is inserted in the device. If the card is inserted incorrectly, the symbol will flash intermittently. Modem – This symbol displays if a modem is connected to the device. (Note: This symbol will vary in appearance depending on the type of modem being used.) 4. Put on your mask assembly when the air starts to flow. 5. Make sure that no air is leaking from your mask into your eyes. If it is, adjust the mask and headgear until the air leak stops. See the instructions provided with your mask for more information. Note: A small amount of mask leak is normal and acceptable. Correct large mask leaks or eye irritation from an air leak as soon as possible.

6. If you are using the device while sleeping, try placing the tubing from the device over your headboard. This may reduce tension on the mask. Note: If you are having trouble with your mask, refer to the instructions included with the mask.

4-2

BIpap auto m SerIeS uSer manual

4.3 uSIng the ramp and BI-flex featureS
You can press the (Ramp) button to activate the Ramp feature, or the Bi-Flex button to change the Bi-Flex setting, if your provider has enabled these features.

4.3.1 ramp feature
The device is equipped with an optional ramp feature that your home care provider can enable or disable. This feature reduces the air pressure when you are trying to fall asleep and then gradually increases (ramps) the pressure until your prescription setting is reached, allowing you to fall asleep more comfortably. If ramp is enabled on your device, after you turn on the airflow, press the button on the top of the device. You can use the Ramp button as often as you wish during the night. Note: If the ramp feature is disabled, nothing will happen when you press the button.

4.3.2 BI-flex Comfort feature
The Bi-Flex comfort feature allows you to adjust the level of air pressure relief that you feel when you exhale during therapy. Your home care provider can enable or disable this feature.

4.3.2.1

BI-flex enaBled

When your provider enables Bi-Flex, a Bi-Flex level will already be set for you on the device. If this is not comfortable for you, you can increase or decrease the setting. To change the Bi-Flex setting, press the Bi-Flex button. The screen shown in Figure 4–4 appears, with an arrow above the current Bi-Flex setting.

Bi-Flex Setting
1 2 3

fIgure 4–4 BI-flex SettIng SCreen The setting of 1 provides a small amount of pressure relief, with higher numbers providing additional relief. To change the pressure relief, press the + or – buttons to increase or decrease the setting. Note: You can also view and change the Bi-Flex setting by navigating to the Bi-Flex Setting screen from the Patient Setup menu (see Section 4.4).

BIpap auto m SerIeS uSer manual

4-3

4.3.2.2

BI-flex loCKed

Your provider may enable Bi-Flex, but lock the setting so you cannot adjust the level. If Bi-Flex is locked, a screen similiar to the one shown in Figure 4-5 appears.

Bi-Flex Setting
1 2 3

fIgure 4–5 BI-flex loCKed SCreen The arrow appears above the setting the provider has prescribed for you, and a lock symbol displays to indicate that you cannot adjust the setting.

4.3.2.3

BI-flex dISaBled

If Bi-Flex has not been prescribed for you and your provider has disabled the Bi-Flex feature, the following screen appears when you press the Bi-Flex button, and you cannot enable or adjust the setting.

Bi-Flex Setting
FLEX comfort has been disabled on this unit.

fIgure 4–6 BI-flex dISaBled SCreen

4.4 ChangIng the devICe SettIngS
You can view the set pressure on the Active Display screen, and you can view the following settings on the Patient Data screens: • • • • • • • • • • • • 4-4 Therapy usage hours Number of sessions greater than 4 hours 90% therapy pressure averages – only appears in Auto Bi-level therapy (if enabled by your home care provider) System leak (if enabled by your home care provider) Apnea/Hypopnea Index (if enabled by your home care provider) Compliance check value Summary Data Bi-Flex (if enabled by your home care provider) Ramp starting pressure (if enabled by your home care provider) Mask alert (enable/disable) Auto off (enable/disable) FOSQ questionnaire BIpap auto m SerIeS uSer manual

Additionally, you can view and modify the following settings on the Patient Setup screens:

4.4.1 navIgatIng the dISplay SCreenS
Use the button to navigate to the next screen and the button to navigate to the previous screen. Use the + and — buttons to adjust the settings on the Patient Setup screens.

4.4.1.1

vIeWIng data on the patIent data SCreenS
Press
– FOSQ Setup +

Figure 4–7 shows how to navigate the Patient Data screens.
Press

Summary Data
Therapy Hours: 0:12 Blower Hours: 0:12 Total Sessions: 0

Therapy Usage
7 Day Avg: 30 Day Avg: Hrs/Night 5:00 8:00

Compliance Check

Sessions > 4
7 Days: 30 Days: 5 Sessions 8 Sessions

3467

Apn-Hyp Index
7 Day Avg: 30 Day Avg: AHI = 0.0 AHI = 0.0

90% Pressure
7 Day Avg: 30 Day Avg: cmH2O 10.0/5.0 12.0/5.0

Note: Only appears if in Auto Bi-level therapy.

System Leak
7 Day Avg: 30 Day Avg:

L/min
0.0 0.0

fIgure 4–7 navIgatIng the patIent data SCreenS Press the button to scroll forward through the data screens, or press the through the screens in the reverse order. The Patient Data screens are described below. 1.
– FOSQ Setup +

button to scroll

Standby Screen The Standby screen appears when you first apply power to the device. You can access the Patient Data menu from this screen by pressing the or buttons. Note: You can access the Patient Setup screens by pressing the + button, and the FOSQ questionnaire by pressing the — button, as described later in this chapter. Note: The FOSQ option only appears if a SmartCard is inserted or a modem is connected to the device.

BIpap auto m SerIeS uSer manual

4-5

Therapy Usage
7 Day Avg: 30 Day Avg: Hrs/Night 5:00 8:00

2.

Therapy Usage Screen This screen displays the number of hours that the device provided therapy, as 7-day and 30-day averages. The maximum value that can be displayed for both averages is 24 hours. Note: This screen is only for reference. Your home care provider may periodically ask you for this information.

Sessions > 4
7 Days: 30 Days: 5 Sessions 8 Sessions

3.

Sessions Greater than 4 Hours Screen This screen displays the number of device therapy sessions that exceeded 4 hours in 7-day and 30-day periods. The maximum value that can be displayed for the 7-day period is 35 sessions, and the maximum value for the 30-day period is 150 sessions. Note: This screen is only for reference. Your home care provider may periodically ask you for this information.

90% Pressure
7 Day Avg: 30 Day Avg: cmH2O 10.0/5.0 12.0/5.0

4.

90% Therapy Pressure Screen This screen displays a 7-day and 30-day average of the 90% pressure that you have used. The 90% pressure is the pressure at which you spent 90% of the session time at or below. For example, if you have the device on (with airflow) for 10 hours, and spend 9 hours at or below 11 cm H2O, and 1 hour above 11 cm H2O, then the 90% pressure is 11 cm H2O. The first number shown is the IPAP average, while the second number shown is the EPAP average. In the example screen at left, the 7-day average is 10.0 cmH2O IPAP, and 5.0 cmH2O EPAP. The maximum value for both the 7-day and 30-day averages is 25.0 cmH2O. Note: This screen only displays if you are in Auto Bi-level therapy and if your home care provider has enabled it. This screen is only for reference. Your home care provider may periodically ask you for this information.

System Leak
7 Day Avg: 30 Day Avg:

5.

L/min
0.0 0.0

System Leak Screen This screen displays a 7-day and 30-day average of the leak history for the device in liters per minute (LPM). System leak is a combination of intentional and unintentional air leak. Intentional leak is the expected leak at the exhalation port. Some leak is required to minimize CO2 rebreathing. Unintentional leak occurs around the patient interface. If you see a large increase in the amount of leak indicated here, contact your home care provider for assistance. Note: This screen only displays if your home care provider has enabled it. It is only for reference. Your home care provider may periodically ask you for this information.

4-6

BIpap auto m SerIeS uSer manual

Apn-Hyp Index
7 Day Avg: 30 Day Avg: AHI = 0.0 AHI = 0.0

6.

Apnea/Hypopnea Index Screen This screen displays your Apnea/Hypopnea index (AHI) for the last 7 and 30 days. Note: This screen only displays if your home care provider has enabled it. It is only for reference. Your home care provider may periodically ask you for this information.

Compliance Check

7.

3467

Compliance Check Value Screen This screen displays a number used by your home care provider to validate that the data provided by you is the data taken from the Summary Data Screen. This is a “read-only” screen and you cannot change the number on the screen. Note: This screen is only for reference. Your home care provider may periodically ask you for this information.

Summary Data
Therapy Hours: 0:12 Blower Hours: 0:12 Total Sessions: 0

8.

Summary Data Screen This screen displays the total therapy hours for the device, the total blower hours, and the total number of sessions greater than 4 hours since the device was last reset by the home care provider. This is a “read-only” screen and you cannot change the numbers on the screen. Note: This screen is only for reference. Your home care provider may periodically ask you for this information.

4.4.1.2

vIeWIng and modIfyIng patIent Setup SCreenS

Figure 4–8 shows how to navigate the Patient Setup screens.

– FOSQ

Setup +

Press +

Bi-Flex Setting
Press

1 2 3

Note: Only appears if Bi-Flex has been enabled Press

Auto Off Feature: ON

Rise Comfort

off 1 2 3

Note: Only appears if Bi-Flex has been disabled

Mask Alert Feature: ON

Ramp Start Pres
4.0 cmH2O

fIgure 4–8 navIgatIng the patIent Setup SCreenS

BIpap auto m SerIeS uSer manual

4-7

The Patient Setup screens are described below. 1.
– FOSQ Setup +

Standby Screen The Standby screen appears when you first apply power to the device. You can access the Patient Setup menu from this screen by pressing the + button. Once inside the Setup menu, press the or buttons to cycle through the screens. Bi-Flex Setting Screen This screen displays the Bi-Flex level set by your home care provider. This setting allows you to adjust the level of air pressure relief that you feel when you exhale during therapy. Your home care provider will either enable, disable, or lock Bi-Flex. – If Bi-Flex is enabled and the setting is not comfortable for you, you can increase or decrease the setting by pressing the + or – buttons. The setting of 1 provides a small amount of pressure relief, with higher numbers providing additional relief. Press the + or – buttons to increase or decrease the setting. – If your provider has locked the Bi-Flex setting, you can view the setting, but cannot change it. The second screen at left shows Bi-Flex locked. Note: This screen only displays if your home care provider has enabled Bi-Flex.

Bi-Flex Setting
1 2 3

2.

Bi-Flex Setting
1 2 3

Rise Comfort

3.

off 1 2 3

Rise Time Comfort Setting Screen Rise time is the time it takes for the device to change from EPAP to IPAP. This screen allows you to adjust the rise time so you can find the desired setting. You can modify the Rise Comfort Setting by 1 when either the + or – button is touched. • 0 (off) reduces the Rise Time feature to the lowest setting (off = 150 msec). • 1 sets Rise Time to 1 (200 msec), and allows the patient to change the setting. • 2 sets Rise Time to 2 (300 msec), and allows the patient to change the setting. • 3 sets Rise Time to 3 (400 msec), and allows the patient to change the setting. This is only available if Bi-Flex has been disabled. Note: This screen will not display if your provider has not enabled Rise Time on your device.

4-8

BIpap auto m SerIeS uSer manual

Ramp Start Pres
4.0 cmH2O

4.

Ramp Starting Pressure Screen You can increase or decrease the ramp starting pressure in 0.5 cm H2O increments by pressing the + or – buttons. The default setting is 4 cm H2O. You can adjust the setting from 4 cm H2O to the prescription pressure setting. Note: This screen will not display if your provider has not enabled Ramp on your device. Mask Alert Screen You can enable or disable the mask alert setting by pressing the + or – buttons to select OFF or ON. If this feature is enabled, the words Mask Leak appear on the Active display screen when a significant mask leak is detected, and an audible alert sounds. Note: If your physician indicates that the Mask Alert is necessary for you, do not disable it. Auto Off Screen You can enable this feature if you want the device to automatically turn the airflow off whenever you remove the interface (e.g., mask) from your airway. Press the + or – buttons to select OFF or ON. If this feature is enabled, the words Auto Off appear on the Standby screen if the device detects that you have removed your mask. The airflow shuts off, and Auto Off continues to appear on the Standby screen until the condition is corrected.

Mask Alert Feature: ON

5.

Auto Off Feature: ON

6.

After you have finished modifying the Patient Setup screens, press the button to return to the Standby screen. Note: The Setup menu automatically times out and returns to the Standby screen after one minute of inactivity.

4.5 CompletIng the foSQ QueStIonnaIre
Note: The FOSQ option only appears on the Standby screen if you have a SmartCard inserted or a modem connected to the device.

From the Standby screen, you can access the FOSQ questionnaire by pressing the – button. The FOSQ test is a ‘quality of life’ questionnaire designed specifically for people with sleep disorders. The results allow health care professionals to see how therapy has improved the quality of your life. By completing the questionnaire periodically, you can provide valuable information about the effectiveness of your treatment. The device can record your answers on the SmartCard or via a wired or wireless modem for later review by the health care professional. Contact your home care provider for further instructions. Note: If your health care professional or home care provider instructs you to complete the questionnaire, he or she will provide you with the questions, and you can enter your answers into the device. Make sure a SmartCard or modem is installed before answering the questions. The FOSQ screens will not display if a SmartCard or modem is not present and able to store data. 4-9

BIpap auto m SerIeS uSer manual

To answer the FOSQ questionnaire, complete the following steps: 1. Make sure that the SmartCard or modem is properly installed. Answers to the questionnaire are saved onto the SmartCard or sent to the health care professional via a wired or wireless modem (when available). If you are using a SmartCard, a SmartCard symbol ( ) appears in the upper right corner of the Active Display screen. If the SmartCard is inserted incorrectly, this symbol flashes to indicate that the card is incorrectly installed, and the FOSQ questionnaire will not be available. 2. From the Standby screen, press the – button to first FOSQ screen, shown in Figure 4–9.

FOSQ 1/30
0 1 2 3 4

fIgure 4–9 foSQ SCreen 1 of 30 Press the + or – button to select your answer to the first question, and an arrow appears over your answer: – 0 indicates that you choose 0 as the answer to the question. – 1 indicates that you choose 1 as the answer to the question. – 2 indicates that you choose 2 as the answer. – 3 indicates that you choose 3 as the answer. – 4 indicates that you choose 4 as the answer. 3. After you answer the first question, press the button to access the next FOSQ screen and continue answering each question. If you need to go back to a previous screen, press the button. 4. There are 30 FOSQ questions. Once you have reached the last question, press the button to save your answers and exit the questionnaire.

4.6 patIent remInder SCreen
Your home care provider may set a patient reminder that will appear periodically to alert you to take certain actions, such as replacing your mask. A message will appear whenever the device transitions from the Active Display to the Standby state if the period of time specified by your provider has expired. You can acknowledge the message by pressing any of the display buttons (+, –, or ). If you do not acknowledge the message within six minutes of it appearing on the screen, it will time out and disappear. The message will continue to display for three consecutive days whenever the device transitions from Active to Standby until it is acknowledged. If, after three days, you still have not acknowledged the message, then the reminder period will reset and no more reminders will display until the next time the patient reminder setting expires. Figure 4–10 shows an example of a patient reminder message.
Patient Reminder

Check your mask, a new one may be available. Call your provider.

fIgure 4–10 patIent remInder SCreen 4-10 BIpap auto m SerIeS uSer manual

Chapter 5: devICe alertS and trouBleShootIng
This chapter describes the device alerts and also provides troubleshooting information for issues you may run into when using the device.

5.1 devICe alertS
The device provides two alert levels, high and medium priority. • High Priority – These alerts require immediate operator response. The alert signal consists of a high priority sound, which is a continuous two-beep pattern (indicated in the following table as: • • • •). Additionally, the backlights on the buttons will provide a high priority flashing pattern consisting of a continuous, bright-to-off, two-flash pattern (indicated in the following table as: ◊◊ ◊◊ ◊◊). Medium Priority – These alerts require prompt operator response. The alert signal consists of a medium priority sound, which is a continuous one-beep pattern (indicated in the following table as: • •). Additionally, the backlights on the buttons will provide a medium priority flashing pattern consisting of a continuous, bright-to-dim, one-flash pattern (indicated in the following table as: ◊ ◊ ◊).



alert Summary taBle
The following table summarizes the high and medium priority alerts. alert
System Error

audIBle IndICator
•• ••

vISual IndICator Backlights: ◊◊ ◊◊ ◊◊ Screen displays the following message: Service Required

devICe aCtIon
The device enters the “Safe state” in which the device power remains on, but the airflow is disabled.

poSSIBle CauSe
Device failure

patIent aCtIon
Press any of the display screen buttons to silence the alert. Remove the power supply cord from the device to remove power. Plug the cord back into the device’s power inlet to restore power. If the alert continues to occur, contact your home care provider.

BIpap auto m SerIeS uSer manual

5-1

alert
SmartCard Error

audIBle IndICator
• •

vISual IndICator
Backlights:

devICe aCtIon
The device continues to operate and provide therapy, but data logging is unavailable and FOSQ is deactivated.

poSSIBle CauSe
A problem exists with the SmartCard inserted in the accessory slot. The card may be improperly inserted or the data may be corrupt.

patIent aCtIon
Remove the SmartCard to clear the alert. Confirm that the card is properly oriented, with the arrow pointing towards the device, as shown below, and reinsert the SmartCard.







Screen displays the following flashing error symbol:

If the alert continues to occur, remove the SmartCard from the device and contact your home care provider. The card may be corrupt and need to be replaced. Mask Leak • • Backlights:







The device continues to operate.

The words Mask Leak flash on the Active Display screen.

The breathing circuit is disconnected or there is a large air leak.

Check your breathing circuit connections and reconnect the tubing if it has come loose. Press any of the display buttons to clear the alarm. Make sure your mask is on properly before you press the button to restart the airflow. If the alert continues to occur, disable the mask alert setting following the instructions in Chapter 4, and continue your therapy overnight. Contact your home care provider the next morning to have your mask checked. You may need a mask refitting.

Auto Off

None

The words Auto Off flash on the Standby screen, indicating that a mask disconnect has been detected.

The airflow shuts off and the device enters the Standby state approximately 45-60 seconds after detection.

The mask has been removed.

Put your mask back on and press the button to turn the airflow on and resume therapy.

5-2

BIpap auto m SerIeS uSer manual

alert
Patient Reminder

audIBle IndICator
None

vISual IndICator
Backlights:

devICe aCtIon
The device continues to operate.

poSSIBle CauSe
Your home care provider may set a patient reminder scheduled to pop up at a particular time to remind you to replace your mask, change your filters, etc.

patIent aCtIon
Press any of the display buttons to acknowledge the message and clear it. If you do not acknowledge the reminder, it will disappear after 6 minutes, and then reappear for three consecutive days whenever the device transitions from the Active state to the Standby state. If you still do not acknowledge the message, the reminder period will reset and the message will not reappear until the next time the Patient Reminder setting expires.







Screen displays a customized reminder message whenever the device transitions from the Active state to the Standby state.

5.2 trouBleShootIng
The table below lists some of the problems you may experience with your device or mask and possible solutions to those problems. proBlem
Nothing happens when you apply power to the device. The audible indicator does not sound and the backlights on the buttons do not light.

Why It happened
There’s no power at the outlet or the device is unplugged.

What to do
If you are using AC power, check the outlet power and verify that the device is properly plugged in. Make sure the AC power cord is connected correctly to the power supply and the power supply cord is securely connected to the device’s power inlet. If the problem continues to occur, contact your home care provider. Return both the device and power supply to your provider, so they can determine if the problem is with the device or power supply. If you are using DC power, make sure your DC power cord and battery adaptor cable connections are secure. Check your battery. It may need recharged or replaced. If the problem persists, check the DC cord’s fuse following the instructions supplied with your DC cord. The fuse may need to be replaced. If the problem still occurs, contact your home care provider. If you are using a humidifier, make sure you follow the instructions for applying power that are provided with your humidifier.

BIpap auto m SerIeS uSer manual

5-3

proBlem
The device does not operate when you press the button. The airflow does not turn on.

Why It happened
There may be a problem with the blower.

What to do
Make sure the device is powered correctly, following the instructions on the previous page. If the audible indicator sounds and the button backlights turn on when you apply power, but the airflow does not turn on, there may be a problem with your device. Contact your home care provider for assistance. Note: When the device is functioning correctly, after you press the button, the device beeps and the airflow turns on after a slight delay. This brief delay is normal.

The device’s display is erratic.

The device or power supply has been dropped or mishandled, or the device or power supply is in an area with high Electromagnetic Interference (EMI) emissions. The device comes installed with troubleshooting software that automatically monitors performance.

Unplug the device and the power supply. Reapply power to the device. If the problem continues, relocate the device to an area with lower EMI emissions (e.g., away from electronic equipment such as cellular phones, cordless phones, computers, TVs, electronic games, hair dryers, etc.). If the problem still occurs, contact your home care provider for assistance.

Device Resets/Reboots: The device shuts down and restarts automatically during therapy. (This is unlikely to occur.)

Such a reset poses no danger to the patient and assures that the patient receives prescribed therapy throughout the night. If there is a possibility of damage to the device, the device will shut down permanently. The product will then display the following system error alert to indicate that the device be returned to the home care provider for service: Service Required

The Ramp feature does not work when you press the Ramp button ( ).

Your home care provider did not prescribe Ramp for you, or your prescription pressure is already set to the minimum setting.

If Ramp has not been prescribed for you, discuss this feature with your home care provider to see if they will change your prescription. If your provider has enabled Ramp, but the feature still does not work, check the pressure setting on your Active Display screen. If it is set to the minimum setting, the Ramp feature will not work.

5-4

BIpap auto m SerIeS uSer manual

proBlem
The device has fallen off your table or nightstand.

Why It happened
The device may not have been properly seated on the nightstand, or the placement of the tubing may have caused the device to fall.

What to do
Always make sure your device is placed on a hard, flat surface so the feet on the bottom of the device can adhere to the surface (make sure there is no fabric under the device). The device (and humidifier, if using) must be level for proper operation. Also, place the device away from the edge of the nightstand or table, so it doesn’t accidentally get knocked off the table. If you are using a humidifier with the device, make sure that the device and humidifier are placed below your head and mask, so that any condensation in the tubing drains back into the water chamber. If the device and humidifier fall and water gets into the device, drain all water out of the device and make sure it is completely dry before reapplying power. If the placement of the tubing causes the device to fall, make sure that you use proper hose management when setting up your device. Route the tubing behind the bed’s headboard, as shown in Chapter 3, Setup. If the device falls or water gets into the device upon falling, let the device dry completely before restarting it. If the device does not operate correctly after falling, contact your home care provider.

You are having problems connecting the tubing to the device.

You have lost the air outlet port or are traveling and forgot to bring the port with you.

If you are not using a humidifier, you cannot connect your tubing to the device without the detachable air outlet port, shown below.

You must have the port attached properly before connecting your tubing. You are experiencing excessive air leaks where the tubing connects to the device. The air outlet port is not installed correctly and doesn’t seal properly. Remove the port and reattach it to make sure it’s properly installed. To remove the port: a. Put your thumb underneath the port and insert your index finger into the port opening. b. Pivoting from the bottom, unsnap the port and pull it away from the device. To properly reattach the port: a. Make sure it is correctly oriented (with the port opening at the top), and insert the two latches at the bottom of the port into the openings on the bottom of the device. b. Push the top of the port down to snap it into place. Reattach your tubing, turn on the airflow, and check to make sure you do not still feel air coming out of the port area.

BIpap auto m SerIeS uSer manual

5-5

proBlem
The air out of the mask is much warmer than usual.

Why It happened
The air filters may be dirty. The device may be operating in direct sunlight or near a heater.

What to do
Clean or replace the air filters as described in Chapter 7. The temperature of the air may vary somewhat based on your room temperature. Make sure that the device is properly ventilated. Keep the device away from bedding or curtains that could block the flow of air around the device. Make sure the device is away from direct sunlight and heating equipment. If the problem continues, contact your home care provider.

The mask feels uncomfortable to wear, there is significant air leakage around the mask, or you experience other maskrelated issues. You have a runny nose.

This could be due to improper headgear adjustment or improper mask fitting, etc. This is caused by a nasal reaction to the airflow. The air is too dry.

If you experience any issues with your mask, refer to your mask instructions for information on proper fitting, etc. If the problem continues, contact your home care provider.

Call your health care professional.

You have throat or nose dryness.

Increase the room humidity. Consult with your health care professional about using a Respironics humidifier with the device. If you have a humidifier, refer to the instructions included with your humidifier to make sure it is working properly. Stop using the device and contact your health care professional.

You experience nasal, sinus, or ear pain.

You may have a sinus or middle ear infection.

5-6

BIpap auto m SerIeS uSer manual

Chapter 6: aCCeSSorIeS
Contact your home care provider for additional information on the accessories available for your BiPAP Auto M Series system. When using optional accessories, always follow the instructions enclosed with the accessories.

6.1 addIng a humIdIfIer
You can use the M Series Heated humidifier or the M Series Pass-over humidifier with your device. They are available from your home care provider. A humidifier may reduce nasal dryness and irritation by adding moisture (and heat, if applicable) to the airflow. Figure 6–1 shows the humidifier by itself and attached to the BiPAP Auto M Series device. Caution: For safe operation, the humidifier must always be positioned below the breathing circuit connection at the mask and the air outlet on the device. The humidifier must be level for proper operation.

Humidifer without Device

Humidifer Device Attached

fIgure 6–1 humIdIfIer alone and attaChed to the devICe Refer to the humidifier’s instructions for complete setup information.

BIpap auto m SerIeS uSer manual

6-1

6.2 uSIng the SmartCard
Some BiPAP Auto M Series systems come with a SmartCard accessory module installed in the accessory slot on the back of the device to record information for the home care provider. Your home care provider may ask you to periodically remove the SmartCard and send it to him/her for evaluation. Note: The SmartCard does not need to be installed for the device to work properly. The SmartCard records device usage information for use by your home care provider. Contact your provider if you have any questions about the SmartCard.

6.3 addIng Supplemental oxygen
Oxygen may be added at the mask connection. Please note the warnings listed below when using oxygen with the device.

WarnIngS
• • When using oxygen with this system, the oxygen supply must comply with local regulations for medical oxygen. When using oxygen with this system, a Respironics Pressure Valve (Part number 302418) must be placed in-line with the patient circuit. Failure to use the pressure valve could result in a fire hazard. Oxygen accelerates fires. Keep the device and the oxygen container away from heat, open flames, any oily substance, or other sources of ignition. Do not smoke in the area near the device or the oxygen. When using oxygen with this system, turn the device on before turning on the oxygen. Turn the oxygen off before turning the device off. This will prevent oxygen accumulation in the device. If administering fixed-flow supplemental oxygen, the oxygen concentration may not be constant. The inspired oxygen concentration will vary, depending on the IPAP and EPAP settings, patient breathing pattern, and leak rate. Substantial leaks around the mask may reduce the inspired oxygen concentration to less than the expected concentrations. Appropriate patient monitoring should be implemented.







6-2

BIpap auto m SerIeS uSer manual

Chapter 7: CleanIng and maIntenanCe
This chapter describes how to clean the device and its filters and provides tips on traveling with your BiPAP Auto M Series system.

7.1 CleanIng the devICe
Warning: Caution: To avoid electrical shock, always unplug the power cord from the wall outlet or DC power source before cleaning the device. Do not immerse the device in liquid or allow any liquid to enter the enclosure, inlet filter, or any opening.

1. Unplug the device, and wipe the outside of the device with a cloth slightly dampened with water and a mild detergent. Let the device dry completely before plugging in the power cord. 2. Inspect the device and all circuit parts for damage after cleaning. Replace any damaged parts.

7.2 CleanIng or replaCIng the fIlterS
Caution: Operating the device with a dirty filter may keep the system from working properly and may damage the device. Under normal usage, you should clean the gray foam filter at least once every two weeks and replace it with a new one every six months. The white ultra-fine filter is disposable and should be replaced after 30 nights of use or sooner if it appears dirty. DO NOT clean the ultra-fine filter. Caution: Dirty inlet filters may cause high operating temperatures that may affect device performance. Regularly examine the inlet filters as needed for integrity and cleanliness. 1. If the device is operating, stop the airflow by pressing the from the power source. button. Disconnect the device

BIpap auto m SerIeS uSer manual

7-1

2. Remove the filter(s) from the enclosure by gently squeezing the filter in the center and pulling it away from the device, as shown in Figure 7–1.

fIgure 7–1 removIng the fIlterS 3. Examine the filter(s) for cleanliness and integrity. 4. Wash the gray foam filter in warm water with a mild detergent. Rinse thoroughly to remove all detergent residue. Allow the filter to air dry completely before reinstalling it. If the foam filter is torn, replace it. (Only Respironics-supplied filters should be used as replacement filters.) 5. If the white ultra-fine filter is dirty or torn, replace it. 6. Reinstall the filters, inserting the white ultra-fine filter first if applicable, as shown in Figure 7–2.
Filter Area

Disposable Ultra-Fine Filter (optional)

Reusable Gray Foam Filter (required)

fIgure 7–2 reInStallIng the fIlterS Caution: Never install a wet filter into the device. It is recommended that you clean the filter in the morning and alternate using the two foam filters provided with the system to ensure sufficient drying time for the cleaned filter.

7-2

BIpap auto m SerIeS uSer manual

7.3 CleanIng the tuBIng
Clean the tubing daily. Disconnect the flexible tubing from the device. Gently wash the tubing in a solution of warm water and a mild detergent. Rinse thoroughly. Air dry.

7.4 ServICe
The BiPAP Auto M Series device does not require routine servicing. Warning: If you notice unexplained changes in the performance of this device, if it is making unusual or harsh sounds, if the device or power supply have been dropped or mishandled, if the enclosure is broken, or if water has entered the device, discontinue use, and contact your home care provider.

7.5 travelIng WIth the SyStem
When traveling, the carrying case is for carry-on luggage only. The carrying case will not protect the system if it is put through checked baggage. For your convenience at security stations, there is a note on the bottom of the device stating that it is medical equipment. It may be helpful to bring this manual along with you to help security personnel understand the BiPAP Auto M Series device. If you typically use a humidifier with your device, but leave the humidifier home when traveling, make sure you attach the air outlet port to your device when you remove the humidifier from the device. You need the port to connect the tubing directly to your device. Figure 7–3 illustrates how to remove the humidifier from the device and attach the air outlet port.

Remove the Device From the Humidifier

Attach Air Outlet Port to the Device

fIgure 7–3 attaChIng the aIr outlet port to the devICe

BIpap auto m SerIeS uSer manual

7-3

7.5.1 InternatIonal travel
If you are traveling to a country with a line voltage different than the one you are currently using, a different power cord or an international plug adaptor may be required to make your power cord compatible with the power outlets of the country to which you are traveling. Contact your home care provider for additional information.

7-4

BIpap auto m SerIeS uSer manual

Chapter 8: SpeCIfICatIonS envIronmental operatIng temperature relatIve humIdIty atmoSpherIC preSSure 41 to 95°F (5 to 35°C) 15 to 95% (non-condensing) 77 to 101kPa (0 to 7500 ft) Storage -4 to 140°F (-20 to 60°C) 15 to 95% (non-condensing) N/A

phySICal
Dimensions: Weight: 7.5” L x 5.0” W x 3.125” H (19 x 12.7 x 7.9 cm) Approximately 2.2 lbs (1 kg) - without a humidifier

StandardS ComplIanCe
This device is designed to conform to the following standards: – – IEC 60601-1 General Requirements for Safety of Medical Electrical Equipment EN ISO 17510-1 Sleep Apnea Breathing Therapy Devices

eleCtrICal
AC Power Consumption: DC Power Consumption: Type of Protection Against Electric Shock: 100 – 240 VAC, 50/60 Hz, 1.0 A max. 12 VDC, 3.0 A max. Class II Equipment

Degree of Protection Against Electric Shock: Type BF Applied Part Degree of Protection against Ingress of Water: Device: Drip Proof, IPX1 Mode of Operation: Electromagnetic Compatibility: Fuses: AC Power Supply: (Reorder number 1015642): Drip Proof, IPX1 Continuous The device meets the requirements of EN 60601-1-2, 2nd edition. There are no user-replaceable fuses.

BIpap auto m SerIeS uSer manual

8-1

preSSure
Pressure Increments: Pressure Stability:

Similar Documents

Free Essay

Cars

...still going to cost you a tidy five-figure sum to buy the necessary parts, materials and paint supplies needed to transform an old car from a worn-out relic to a dazzling show beauty. If you've decided to restore a particular old car because you have a longstanding attachment to that model and are not concerned about its value or whether parts can be easily found, then we applaud you--that mindset should be the primary focus of this great hobby of ours. However, if you enjoy old car restoration but prefer your projects to be easy and relatively cheap to do--with, perhaps, the added bonus of making a few dollars on it when it's done--then these are the cars you should consider for your next venture. Of course, the majority of these cars won't make you stand apart from the crowd at a show or cruise-in, but you will be able to complete their restoration far faster and a whole lot cheaper than if you chose a car that was rare, or had only been made for a few years. Another factor to consider is the actual purchase price of the car. While some of the cars listed here can be bought for around $5,000, others will cost four times as much even if they're basket cases--but keep in mind that those higher-priced cars will also bring four times the money when sold, so it kind of equals out. But the one common feature enjoyed by most of these cars--and the main reason why they are on this list--is that they're backed by a huge parts supply, making them a lot easier to restore...

Words: 485 - Pages: 2

Free Essay

Cars

... What I find to be awesome is cars. Cars are the best thing to ever happen to this world. Where would we be without cars? Cars are the signature to most advancement. Cars are fast. Cars are classic. Cars are just amazing. I’m sure most people don’t ever stop to admire a car. Cars are amazing if you actually know about them and know what you like. There is amazing cars all over the world I have lots of cars I really like. Cars are an art. Nothing is better than the bond you can make by building your own car. It’s a whole other feeling. Cars are amazing. For three reasons. First reason is that they are fast. In my opinion of a good fast car. I would get a 1967 Ford Mustang Shelby GT500. Just because that’s my style. This car is rolling fast yea there is faster imports, but this is classic, friendly but crazy fast. This is a beast on 4 wheels. 271 horse power you bet its fast. This car goes from 0-60 in 4.3 seconds. This is just one of the millions of cars out there. Cars are legit the best past time because of their incredible speed. Another fast car is the 1987 Grand National. This car is fast and slick, you need a getaway car this is your car. Its precise steering with the right driver you can drive miracles. It goes from 0-60 in 4.8 seconds. No matter what car you have you get or are thinking of getting how fast is it or how fast does it go is always one of the questions you ask. Okay so the second reason I think Cars are amazing is because they are...

Words: 676 - Pages: 3

Premium Essay

Cars

...California, a Prius—one of the cars in Google’s fleet of autonomous vehicles—caused a fender-bender. The accident happened, naturally, because there was a human behind the wheel. That's according to a new report by Google about its self-driving car project, a document that tallies every accident since it began testing its 32 vehicles on the roads in 2009. That sounds like it could be a massive dataset, considering the tens of thousands of accidents that happen on American roadways each year. It actually isn’t: “In the six years of our project, we’ve been involved in 12 minor accidents during more than 1.8 million miles of autonomous and manual driving combined,” Google wrote. “Not once was the self-driving car the cause of the accident.” Not once! In the case of the fender bender four summers ago, a Google employee—yes, a human—was to blame. He had borrowed the car to run a quick errand and ended up rear-ending another car. (“He was not using the vehicle to test our autonomous technology,” Google wrote, and the car ended up sustaining “some damage.”) So what ended up being probably a pretty bad day for that Google employee—imagine telling your boss you crashed the driverless car—turns out to be a tidy example of the key difference between human drivers and algorithmic ones. Namely: Robots are much, much better drivers than humans. Google’s descriptions of the other accidents involving its self-driving cars reiterates that point. Its cars were rear-ended eight times,...

Words: 806 - Pages: 4

Premium Essay

Driverless Cars

...Google Cars Darrell Jackson Jr. Composition II Cars that can drive themselves are no longer a part of science fiction. Google is currently in the process of changing the world as we know it. Google is working on a project called “driverless cars”. “What is a driverless car” you ask? A driverless car is an autonomous vehicle that can drive itself from one place to another without the assistance from a human driver. The driverless car is a project by Google that enables your car to drive without really driving. Although these cars have yet to hit the roads, they are already being tested by Google and many different car companies. Google has complete over 700,000 driving miles accident free, in which they traveled congested streets and popular intersections. According to predictions driverless cars could be a reality by 2020. Google is still in the beginning stages of making this project official because there is a lot of barriers that have to get through, but confidently state driverless cars will become available in the near future. Google have at least 10 different prototypes that they use as practice cars. Some examples of these cars are the Toyota Prius and the Lexus SL600h. The basic designs of this project consist of a car with a rotating sensor on the top that spins 360 degrees, to create a 3D map of the cars location. A sensor located on the left rear wheel that tracks the car movements and locates it on the map. There is a video camera on the windshield by the rear...

Words: 1457 - Pages: 6

Premium Essay

Nano Cars

...Mr. J. Mateo Student Name/ID: Sony Surendran 09093103 PRODUCT DEVELOPMENT AND STRATEGICS Course work 1 Consider any consumer product that you think needs improvement. a. Why is improvement necessary? b. What improvement do you think, should be made on the product? c. How will these improvements affect the marketability of the product? Product: TATA NANO: The world’s cheapest 4-door hatch back car. Introduction Tata Nano is a rear-engine fixed city car able to travel four adults, built by Tata Motors(India). Targeting primarily for the Indian market. The car has a fuel economy of around 26 kilometers per liter of petrol on the highway and around 23 kilometers per liter in the city riding conditions. It announced at the 9th annual Auto Expo on January 10, 2008, in New Delhi, India by TATA’s Managing director RATHAN TATA. The Nano had its commercial launch on March 23, 2009 and a booking period from April 9 to April 25, generating more than 200,000 bookings for the car. The Nano cars started sales to Indian customers after July 17, 2009 with a starting price of Rs 100,000, which is approximately equal to S$ 3,125 or US$2,232 as of March-21 2010. The dual cylinder 624cc petrol engine is an all aluminium unit and is rated at 35 BHP (5,250 rpm) with 48 NM of torque (3,000 rpm). The rear mounted engine transfers power directly to the rear...

Words: 257 - Pages: 2

Free Essay

Cars

...2017 Lincoln Town Car design 2017 Lincoln Town Car is a luxurious sedan car designed to give full luxury. Crown Victoria model is said to share its architectural design with is this new paradigm, 2017 Lincoln Town Car design. Therefore, Lincoln will be unveiling this new brand model with some predecessors styles element found in it though the design and concepts will be new. 2017 Lincoln Town Car will have a 400-HP version of the 3.0-liter supercharged Eco Boost motor with a concise designation 3.0 T modifications. Its body build from aluminum creates comfy drive as this will reduce its weight. The car has soft pneumatic suspension combined with sound insulation. The significant difference can be noted from its large wing grille and LED headlights. The chassis will be enhanced compared to the previous models. From its weight, we will have a better performance of the car with low fuel consumption. Its interior is incredibly fantastic as it is furnished with leather seats and preinstalled infotainment systems as well as satellite navigation. The suspension system is also upgraded for a more comfy zone without forgetting the consoles which have been upgraded too. The car will have a single choice for the engine as it is 2.7 l twin-turbocharged V6 EcoBoost. The incredible part comes from its 350HP more than 370 lb-ft of torque. The gearbox will be of automatic of 9-speed. Its acceleration is fantastic as one will be able to acceleration from 0 to 60mph in six seconds accelerating...

Words: 401 - Pages: 2

Free Essay

Dream Car

...My dream car would be a 1969 Chevy Camaro with a 350 engine and a four speed transmission. The name "Camaro" came from a French word for friend. The decision on the name came down to the last minute, with most of the world sure the car would be named, "The Panther." But the name Camaro fit in with other Chevy names- Corvette, Chevelle, Chevy 2, and Corvair. The main reason the Camaro was introduced was because of the huge success of the Ford Mustang. The Camaro was roughly the same size as the Mustang, a little wider and based more on high performance. The Camaro is one of the last remaining muscle cars still in production today. It is only fitting that buyers still expect maximum performance from their Camaros. It is that expectation that has kept the Camaro alive for all these years, while many other cars have faded away, lost in memory. The Mustang GT only offered the 289- cubic inch or a 390- cubic inch V8 in 1967. The Camaro rolled out with 302, 327, 350, and 396 cubic inch V8's The Camaro's style was much smoother as well. The introduction of the Camaro threw pony car (Mustang) development into frenzy. Before the Camaro, the Mustang and Barracuda were not quite considered full muscle cars. Most serious performance enthusiasts still opted for intermediate sized GTO's or the Chevelle Super Sports (SS). The Camaro changed the image of those sport coupes. The Z28 and the stout SS-396 were more than just a stylish ride. Under the hood Camaros were well respected...

Words: 484 - Pages: 2

Premium Essay

Google Cars

...GOOGLE SELF-DRIVEN CAR THESIS NAME: INSTITUTION: COURSE PREFIX: COURSE NUMBER: DATE OF SUBMISSION: 1. INTRODUCTION 1.1 GOOGLE SELF-DRIVEN CAR The Google self-driven car is an automated car that operates automatically with no physical human input. It is a project developed by Google Company that involves the development of technology in automated vehicles mostly the electric type of vehicles. The software responsible for the management of the vehicle in its operations is the Google chauffeur. The project is under the management of a Google engineer Sebastian Thrun. He is an experienced engineer who previously worked for the former director in the Stanford Artificial Intelligence Laboratory. In addition, he previously worked as a co-inventor of the Google Street view. The development of the robotic vehicle was first showcased in 2005 in the DARPA Grand Challenge where it won the award. There was an able team of 15 engineers who developed the system in collaboration with the Google Company. Some included Chris Urmson, Mike Montemerlo and lastly Anthony Levandowski, involved in the DARPA Grand and Urban Challenges (Thrun, 2010). In the technological point of view, the automated Google vehicles comprise of parts and equipment that are worth 150000 dollars inclusive of the 70000 dollars for the LiDAR system. It possesses a range finder strategically placed on the top part of the vehicle called the Velodyne...

Words: 5459 - Pages: 22

Premium Essay

Elecric Cars

...News regarding electric cars has been rampant for the past year or so, and for good reason. New models are out, and these cars can help the country lose some of its dependency on oil while providing us with an alternative that is gentler on the environment. These latest-generation electric cars have better range and performance than ever before. Electric cars are the greenest of the personal transportation industry. They do not emit any (or much less) carbon emissions than their usual counterparts, but they do utilize questionable raw materials such as the metals that go into their batteries. There are some major advantages of having an electric car to the environmental but there are also some disadvantages. It is important to know all of them before making a purchase. The greatest advantage of an electric car is obviously the complete lack of emissions because they run on battery power. Emissions are the greatest pollutant of the environment as they cause instability in the atmosphere. Some vehicles like the Chevy Volt, only run on batter power for the first forty or so miles of driving then start being powered by gasoline. The new 2011 Nissan Leaf or Tesla Roadster are full electric vehicle however, will not even have an exhaust system, muffler or fuel tank. They do not use any fossil fuels internally. They are fueled completely by a rechargeable battery that does not have to be charged via a gas engine. With many hybrid cars, there is a gas engine to go along with the electric...

Words: 335 - Pages: 2

Premium Essay

Driverless Cars

...so farfetched. Driverless cars are fast becoming reality. Just as computers did, driverless cars will change the way we live. And like any other innovation, it would free up time for more creative thinking. In recent years, we have seen small improvements to our cars, making them more automated. Bluetooth and GPS play a big role while driving our cars. GPS will navigate for us without using a map, and Bluetooth allows us to go hands free to concentrate on driving. So, is a fully autonomous car good or bad? Google is leading the way trying to make Autonomous vehicles a part of people’s everyday lives. Other car manufactures are making an effort to make driverless vehicles, but, google is by far is ahead of the pact. According to the article How Will Driverless Cars Affect Our Cities? “Google is the most conspicuous developer of autonomous vehicles, but it is hardly alone in pursuing this venture”. (ISSA ROMAM, 2013) Cars without drivers could possibly solve some of our more common issues on the road such as traffic congestion, accidents, and human error. According to the article Driverless Cars Switch into Overdrive, “Google’s driverless cars have logged 50,000 miles on the road without help from a driver. The Google cars can maintain safe driving distances in moving traffic, park, and brake to avoid collisions. They have managed to log all of those miles without any accidents”. (Elise Grafe, 2013) In the article How Will Driverless Cars Affect Our Cities Issa Roma...

Words: 789 - Pages: 4

Premium Essay

Google Car

...Google, want to venture into the self-driving cars, they announced that it is building a vehicle without a steering wheel or role for a driver. The thing is that Google technology has not been able to successfully switch control back and forth from automated driving to the driver in an emergency and does not expect to be able to accomplish that soon, also technology is to expensive. Other companies building autonomous cars said that they are working on vehicles that will be able to safely make that switch. Volvo says that it expects to have its cars tested on city streets by 2020. The Google management team faced several questions. Should Google continue to invest in the technology behind self-driving cars? How could Google's core software-based and search business benefit from self-driving car technology? As large auto manufacturers began to invest in automotive technology themselves, could Google compete? Was this investment of time and resources worth it for Google? Environmental Scanning (external and internal analysis): PEST Analysis After reading the case study Google car I have used the PEST analysis model in which the following was concluded: 1. Political Factors: Government’s regulations on automobile transportation are one of the major aspects in Google's car that have legal barriers. Due to that most of the governments do not have specific laws for self-driving vehicle and there are no federal standards for autonomous cars special permission in all US states adjusted...

Words: 479 - Pages: 2

Free Essay

Muscle Cars

...Muscle Cars It has been said that racing began once the second car was built. NASCAR started the “race” of the American muscle car in 1947. Due to the growing interest in power and speed, General Motors introduced the 1949 Oldsmobile Rocket 88. America’s first high compression overhead-valve V8 was put into the light body of the Oldsmobile, creating the first muscle car. American’s take pride in owning the biggest, most eccentric vehicles. Due to the gas prices in other countries such as Europe and Asia, they are less likely able to afford these large, gas thirsty vehicles. Americans have been infatuated with making cars faster and even more powerful since the Rocket 88, therefore, the muscle car is a huge symbol of American culture. Muscle cars were being produced by many different manufacturers in the early 1950’s, including Chevrolet, Chrysler, Dodge, Ford and Pontiac. In the 1950’s bigger was considered better. These muscle cars gained weight and inches with every year that past. Gaining weight meant losing speed, the cars required larger engines, but still weren’t fast enough for the buying market. Each manufacturer had a different idea to enhance performance and boost power. Chevy only offered fuel injection, Pontiac offered fuel injection and “tri-power”, and Ford offered supercharging. Dodge went in at a different angle and offered a power up “D-500” engine for every 1956 model they had. In 1955, Chrysler C-300 AKA “America’s Most Powerful Car” became the star of NASCAR...

Words: 642 - Pages: 3

Premium Essay

Electric Cars

...It is a necessity to own or to have access to a car in order to keep up with all the competition of the business world, and also one’s social demands. Most people would not be able to travel around a country or the world without this incredible machine, for it provides freedom and mobility, even for people who do not own a car. As technology has advanced so has the automobile. Unfortunately the car has a very destructive nature putting out a lot of dirty emissions into our atmosphere. That’s why more and more people are turning to cleaner ways of getting around by choosing the electric car as an option. However even though electric cars sound like the best deal in today’s world the technology is not there yet for them to push out gasoline powered cars out of the way, in other words, gasoline powered automobiles should not be pushed towards being electric powered. The electricity and elements charging the electric car would have to come from nuclear and coal power plants, the electric car is not dependable for long trips and they are priced, on average, much higher than our gasoline powered vehicles. All electric cars are zero emission vehicles. It is clear that vehicle emissions wreak havoc on air quality, and it’s particularly evident in big cities like Boston, New York and Los Angeles. “Clean vehicle and improved fuel technologies can significantly reduce air pollution. One way to address the pollution caused by gas-fueled cars is to move toward electrically powered automobiles...

Words: 946 - Pages: 4

Premium Essay

Electirc Car

...of the Electric Car 1832-1839 Scottish inventor Robert Anderson invents the first crude electric carriage powered by non-rechargeable primary cells. 1835 American Thomas Davenport is credited with building the first practicalelectric vehicle -- a small locomotive. 1859 French physicist Gaston Planté invents the rechargeable lead-acid storage battery. In 1881, his countryman Camille Faure will improve the storage battery's ability to supply current and invent the basic lead-acid battery used in automobiles. 1891 William Morrison of Des Moines, Iowa builds the first successful electric automobile in the United States. Thomas Edison and an electric car. Courtesy of the Smithsonian 1893 A handful of different makes and models of electric cars are exhibited in Chicago. 1897 The first electric taxis hit the streets of New York City early in the year. The Pope Manufacturing Company of Connecticut becomes the first large-scale American electric automobile manufacturer. 1899 Believing that electricity will run autos in the future, Thomas Alva Edison begins his mission to create a long-lasting, powerful battery for commercial automobiles. Though his research yields some improvements to the alkaline battery, he ultimately abandons his quest a decade later. 1900 The electric automobile is in its heyday. Of the 4,192 cars produced in the United States 28 percent are powered by electricity, and electric autos represent about one-third of all cars found on the roads...

Words: 1824 - Pages: 8

Free Essay

Street Car

...but it depends on the market they are targeting on. Like in Pakistan most of the people are middle class families who cannot afford to drive a car so for them this scheme does not create any difference for them. Secondly, it would be an expensive deal for Pakistani market because as per petrol prices are rising day by day and as the crime rate are increasing and car snatching is so common that no company would take risk to invest in Pakistani market. 2. How would this idea compete with the traditional car rental industry? Unlike traditional car rental industries, it eliminates the long lines and frustration at the rental-car office, and gives its members the grab-and-go convenience of renting a vehicle 24 hours a day, paying only for only as long as needed, and usually with little advance reservation notice. And with no fuel, insurance or maintenance costs, Street Car has become a great option for drivers with minimal transportation needs, and those with no interest in the price and parking headaches of vehicle ownership. It also provides ease and convenience, which are the greatest advantages of car sharing, behind the obvious financial savings. Unlike a traditional rental-car office, Streetcars clients climb behind the wheel and drive off the parking lot in seconds with no lines, no waits, no paperwork. Members reserve their car by computer, phone, text message or cellular app. Simply fill in the preference blanks - what time, how long, pickup point, type of vehicle -...

Words: 348 - Pages: 2