Free Essay

Uv Calibration

In:

Submitted By Siddheshswami
Words 982
Pages 4
UV-Visible Spectrophotometer
Calibration
S.ZALI
M.Sc.
Dept. of Analytical chemistry
Razi vaccine serum research institute
S.Zali@rvsri.ir
2008

The primary objective of GLP is to ensure the generation of high quality data. Essential to this is the ability to produce estimates for the measurement uncertainty associated with calibrations.

V

The spectrophotometer has become the workhorse for high throughput routine testing for a wide variety of analysts in chemical and biochemical analysis laboratories. It is important for accuracy and reproducibility in measurement that this instrument is correctly set up and calibrated.

Increasing Wavelength
50,000

Gamma

X-ray

12,820

UV
200

4000

NIR
380

780

IR
2500

400 cm-1

FIR

Micro

Radio

25 000 nm

Increasing Energy

K-shell electrons Outer-shell electrons Molecular vibrations Molecular
Rotation

Conventional Spectrophotometer:

The factors that must be calibrated:
Control of wavelengths
Control of absorbance
Limit of stray light
Resolution
Spectral slit width
Cell

Control of wavelengths:
Wavelength standards:
The ideal wavelength standard would have very narrow and very well defined peaks.

Control of wavelengths:
•Solution of holmium(III) in perchloric acid: 240–
640 nm.
•Holmium oxide filters: 280–640 nm.
•Didymium filters: 400–750 nm.
•Using of spectral lines in low-pressure discharge lamps: 656.1 nm produced by the D2 lamp.

Holmium(III) in perchloric acid:
Consists of a 4% W/V solution of Holmium Oxide in
1.4M Perchloric Acid. λmax………………241.1 ± 0.4nm
278.0 ± 0.4nm
287.5 ± 0.4nm
361.2 ± 0.4nm
416.6 ± 0.4nm
451.3 ± 0.4nm
458.3 ± 0.4nm
537.5 ± 0.4nm
640.8 ± 0.4nm

Holmium oxide filters:

Didymium filters:

Discharge lamps:
A quick check of the wavelength accuracy of a spectrophotometer can be performed by observing the strong emission band at 656.1 nm or 486.0nm produced by the D2 lamp.
In fact many of modern, so-called self calibrating instruments make use of deuterium emission lines to set the wavelength scales.

British pharmacopoeia 2007:



The permitted tolerance is ± 1 nm for the ultraviolet range and ± 3 nm for the visible range. Suitable certified reference materials may also be used.

United Stated pharmacopoeia:
The 486.13 - nm and 656.28-nm lines of a hydrogen discharge lamp.
Suitable glass filters.
Standard glasses containing didymium.
Standard glasses containing holmium.
Standard holmium oxide solution.

Control of absorbance:
In the visible by convention measurements are often made at the wavelength of the 546-nm mercury line.
The Neutral density glasses for 400–1,000 nm.
Copper sulphate in H2SO4 (0.1 mol/1).
Cobalt ammonium sulphate in H2SO4 (0.1 mol/1).
Potassium dichromate in H2SO4 (0.005 mol/1).
Potassium dichromate in perchloric acid (0.001 mol/1).

British pharmacopoeia 2007:


Check the absorbance using suitable filters or a solution of potassium dichromate R at the wavelengths indicated in Table 2.2.25.-2



The tolerance for the absorbance is ± 0.01.

United Stated pharmacopoeia:
USP

USP

A number of standard inorganic glass filters as well as standard solutions of known transmittances such as potassium dichromate are available.
USP
USP

USP

Limit of stray light:
Heterochromatic stray light

Isochromatic stray light

Heterochromatic stray light:
1. Liquid filters:
– KCl aqueous (12 g/1) at 175–200 nm
– NaI aqueous (10 g/1) at 210–259 nm
– Acetone at 250–320 nm
– NaNO2 aqueous (50 g/1) at 300–385 nm
2. Sharp-cut filter glasses.

Isochromatic stray light:

British pharmacopoeia 2007:
Stray light may be detected at a given wavelength with suitable filters or solutions: for example the absorbance of a 12 g/l solution of potassium chloride
R in a 1 cm cell increases steeply between 220 nm and 200 nm and is greater than 2.0 at 198 nm when compared with water as compensation liquid.
Suitable certified reference materials may also be used. Control of cuvettes: measure the transmittance of water-filled cuvettes against air
– For glass cells at 356 and 650 nm
– For quartz cells at 220 and 240 nm
0.85 < at the lower wavelength
0.88 < at the longer wavelength

Huber L (2002) Validation and qualification in analytical laboratories. Interpharm Press, Denver, USA

British pharmacopoeia 2007:
The tolerance on the path length of the cells used is ±
0.005cm.
When filled with the same solvent, the cells intended to contain the solution to be examined and the compensation liquid must have the same transmittance. Wavelength resolution:
The operational qualification of the instrument should then include a test in which the spectra absorbance of a solution of toluene in hexane (0.2 mol l-1) is measured and the ratio of the absorbance at the 266-nm maximum to that at the 269-nm minimum is determined. This peak-to-valley ratio should be greater than 1.5.

British pharmacopoeia 2007:
Record the spectrum of a 0.02 percent V/V solution of toluene R in hexane R. The minimum ratio of the absorbance at the maximum at 269 nm to that at the minimum at 266 nm is stated in the monograph. Suitable certified reference materials may also be used. Spectral slit-width
(for quantitative analysis):
To avoid errors due to spectral slit-width, when using an instrument on which the slit-width is variable at the selected wavelength, the slit-width must be small compared with the half-width of the absorption band but it must be as large as possible to obtain a high value of I0. Therefore, a slit-width is chosen such that further reduction does not result in a change in absorbance reading.

References:
•British Pharmacopoeia 2007
•US Pharmacopoeia 30
•Calibration of UV–vis spectrophotometers for chemical analysis
/Accred Qual Assur (2005) /John F. Clare
•UV Spectroscopy Techniques, instrumentation,data handling./B.J.Clark, T.Frost and M.A.Russell
•Standards in absorption spectrometry/C.Bugess and A.Knowles

Questions

Thank you for your attention

Similar Documents

Free Essay

Sdfb

...THE POTENTIOMETER: Internal Resistance of a Test Cell OBJECT To calibrate a one meter slide wire potentiometer using a standard cell and then to use this potentiometer to measure the emf of a test cell. The terminal voltage of the same test cell is then measured as different load resistors are connected across the test cell and these data are used to determine the internal resistance of the test cell. THEORY The electromotive force (emf) of a cell is its terminal voltage when no current is flowing through it. The terminal voltage of a cell is the potential difference between its electrodes. A voltmeter cannot be used to measure the emf of a cell because a voltmeter draws some current from the cell. To measure a cell's emf a potentiometer is used since in a potentiometer measurement no current is flowing. It employs a null method of measuring potential difference, so that when a balance is reached and the reading is being taken, no current is drawn from the source to be measured. Figure 1. This is the basic circuit diagram for a potentiometer.  Point C is the sliding contact which can be adjusted for zero current deflection through the galvanometer. In this method (refer to Figure 1) a uniform, bare slide wire AB is connected across the power supply. If you were to connect a voltmeter between the + power supply terminal and point A you would measure essentially zero volts. If you were to now connect the voltmeter between the + power supply and point B you would measure...

Words: 1527 - Pages: 7

Premium Essay

Examples of Possible Exam Questions Sustainable Business and Ethics Course

...We expect that you would be able to find at least 10 to 15 standards for your product. For each standard, you should be able to indicate the answers to each column of the CSR standards inventarisation excel sheet. Namely: Name of the standards you have found relevant to your chosen product Explanation of the purpose of the standard Applicable countries Applicable trade region(s) Application to the chosen product Website address(es) Possible market consequences for the supply chain Example: Name of the standards you have found relevant to your chosen product: Florverde Explanation of the purpose of the standard: Florverde is an initiative of ASOCOFLORES the Colombian association of flowers exporters. The aim of the combined action is to satisfy the requirements of Colombian’s flower 1st importer market the USA. The Florverde certification set up in 1996 enables the thousands of workers involved in the production of export flowers to live and work in acceptable conditions. The certification secures the main USA market and maintain best sustainable practices through annual audits. The main areas for standards include: basic labor regulations, personnel management, work and social welfare, water and irrigation, intergrated pest and disease management, traceability and records. Applicable countries: Colombia Applicable trade region(s): CAN (Andean community) Application to the chosen product: 1. Describe in a much details possible the specifications...

Words: 456 - Pages: 2

Free Essay

Zghgfhgf

...Math – Final Exam Vet 110 Name: _______________________ Date: _______________________ Good luck! ( Identify the highest value: 1) 3.7 4.1 2.9 2) 5.4 6.2 4.7 3) 1.2 1.5 1.7 Add the following decimals: 4) 1.35 + 2.42 = ________ 5) 2.8 + 0.03 = ________ Subtract the following decimals: 6) 2.3 – 1.45 = ________ 7) 10.25 – 1.47 = ________ Multiply the following: 8) 1.37 x 0.02 = ________ 9) 0.31 x 0.04 = ________ Eliminate the decimal: 10) 5.25 = ________ 0.3 Solve the following equations; express your answer to the nearest tenth. 11) 12 x 500 x 2000 x 1 = ________ 1 2700 1 60 12) 0.35 x 2.5 = ________ 7. 0.7 13) How many tablets will be needed to give the following dose? The tablets are labeled 1.2 mg. The order is for 0.6 mg. ________ tablet(s) 14) Reduce the following as far as possible: 124 76 = ___________________________ 15) Express the following division to the nearest tenth: 5.7 1.46 = __________________________ 16) Solve the following equation. Express answer to the nearest tenth. 30 x 50 x 10 x 415 1 550 1250 1 = ____________________ Write the following metric measures using their abbreviations and notation rules. 17) Four hundredth of a gram ___________ 18) ...

Words: 648 - Pages: 3

Free Essay

My Test

...| | | | | | | |INSPECTION / |STANDARD (1) |PROCEDURE AND/OR ACCEPTANCE CRITERIA |DOC |SAMPLE |CUSTOMER | |TEST ACTIVITY | | |Y/N(2) |SIZE |INSPECTION | |OPERATION ACTIVITY: 0010 |IN PROCESS | | | |VERIFY PART(S) HAVE RAW MATERIAL TRACEABILITY IS AVAILABLE BEFORE USE | | | | |DRSV010 |CAMERON STANDARD |METHOD- QUALIFIED PERSONNEL TO VERIFY APPLICABLE MATERIAL IS CORRECT BEFORE USE |YES |100% |- | |RAW MATERIAL | |STANDARD- CAMERON STANDARD | | ...

Words: 475 - Pages: 2

Free Essay

Quality One Call at a Time

...RUNNING HEAD: Quality One Call at a Time Quality One Call at a Time May 30, 2006 Quality scores in many call centers are on the decline. The main goal of a call center is to provide excellent customer service while delivering accurate information in a timely manner. “Fast is fine, but accuracy is everything.” Wyatt Earp. Wyatt Earp may not have said this quote in reference to managing quality at call centers, but it carries the same relevance. Management at call centers tends to focus on other issues instead of the accuracy of the information that the CSR is relaying to the customer. Thus value is misplaced and the quality at call centers starts to head on a downward path. One issue that could drive quality scores down is ineffective training. At a call center here in Jacksonville, management decided to pull CSRs out of training after the second day to answer calls because the queue was high. This interrupted the training process and added days to the training schedule. The trainer manager was powerless to prevent this from happening. This happened 6 times over the course of the proposed three-week training calendar. Call center trainers must ensure that their trainees’ learning is not hindered by management’s goal to reduce queue times. The constant yo-yo handling of the CSRs resulted in training not being delivered on all topics, which resulted in low quality scores. This has not been a high topic of focus for most call center management staff. ...

Words: 836 - Pages: 4

Premium Essay

Surface Well Test

...SURFACE WELL TESTING By Karrym Rebay Vol. 1 Preface This is a paper that has been produced to help you better understand and work with the equipment you are familiar with and presently operating on your particular Well Test Unit. It will give you more knowledge about, not only the operation, but also the maintenance and calibration techniques that will help you become more familiar with your own equipment. Paper Content 1. PAPER DESCRIPTION 2. IDENTIFYING A STANDARD WELLTEST PACKAGE AND ITS COMPONENTS 3. EQUIPMENT OPERATION AND FLOWING CONDITIONS 4. SAFETY AROUND YOUR EQUIPMENT 5. CALIBRATION AND MAINTENANCE 6. CALCULATING FLOW RATES 7. GLOSSARY – Exploration & Production Terms 8. CONVERSIONS & TABLES Paper Description This paper describes the specific conditions under which well tests must be performed, lists the surface testing equipment used to perform these well tests, summarizes how this equipment is used to collect samples at the surface and lists several examples that influence the layout of surface equipment. A reservoir test can only be performed under certain conditions. This means the reservoir must be exposed to a disturbance that will cause the reservoir pressure to change. This pressure change, when recorded and interpreted along with the measured flow rates, will give us information about well and reservoir parameters and geometry. A pressure disturbance...

Words: 873 - Pages: 4

Free Essay

Joel Hills Paper

...Wilson Electronics Cellular Booster Database Requirements For Cellular Booster O-14828 Version 1.06 Presented to: Wilson Electronics Approval Name Title Signature Date Confidential Wilson Electronics Cellular Booster Database Requirements - v1.06 O-14828 Change History Version # 1.00 1.01 1.02 1.03 Reason Author Date 2014/04/11 2014/04/16 2014/04/21 2014/04/22 Initial Release Condensed parameters Consensus reached on majority of parameters Added specifics about DB connection, changed AGC Test Parameters Moved some configuration items to test limit sections Al Braga Al Braga Al Braga Al Braga Al Braga 1.04 2014/04/30 1.05 1.06 Merged RF Test Sections, added error, clarified Al Braga some parameter types. Removed all Station Name inputs Al Braga 2014/05/14 2014/05/15 Page 2 of 14 Project File: Cellular Booster Database Requirements.docx ©Averna - Confidential Wilson Electronics Cellular Booster Database Requirements - v1.06 O-14828 Table of content Purpose and Scope ....................................................................................................................................... 4 Considerations .............................................................................................................................................. 4 Database Connectivity .................................................................................................................................. 5 Database...

Words: 2582 - Pages: 11

Premium Essay

Quantitative Analysis of an Unknown Liquid Sample

... Quantitative Analysis of an Unknown Liquid Sample Objective: To be familiar with Ultra Violet Spectrophotometer (UV) and Gas Chromatography (GC) used in chemical analysis UV Abstract In this report, people who consumed tom yum soup suffered from nausea and vomiting due to copper poisoning which was found to leach from a pot into the soup under high heat and acidic condition. It was also suspected that an organic compound was present in the soup which enhanced the absorption of copper in consumer. Hence, as an analytical chemist, we have to use UV to determine the actual concentration of copper standards and blank using external calibration standards. The result of the test solution was measured by comparing it with the calibration of copper. Introduction The goal of this experiment is to obtain the concentration of copper in a known solution. Therefore, in this experiment, we will be using UV to measure the absorbance of the solution. A spectrophotometer is used to measure the amount of light that a sample absorbs. Ultraviolet (UV) light is an electromagnetic radiation with a wavelength in the range of 10nm to 400nm and energy from 3eV to 124Ev. With a shorter wavelength as compared to visible light, UV light is able to penetrate more readily through obstacles. Its name came from a spectrum which humans identify as the colour violet. UV light is invisible, but it can be seen indirectly when it makes other substances glows in visible spectrum. Applications ...

Words: 2224 - Pages: 9

Premium Essay

Nt1310 Unit 9 Lab Report

...Preformulation studies: Identification of drug was carried out by DSC, FTIR and UV spectrophotometry 4A.1.1 DSC studies-DSC studies were carried out at a temperature range of 25 º C to 165 º C at 10 º C increments of temperature. Figure4A.1-DSC Thermogram of Aceclofenac  DSC thermogram indicates the M.P of 154º C which fairly matches with reported M.P of Aceclofenac, thus confirming the identity and purity of aceclofenac. 4A.1.2. FTIR studies- Figure4A.2 . FTIR spectra of Aceclofenac Figure.4A.3 Standard FTIR spectra of aceclofenac FTIR Interpretation: Figure 4A.4 Chemical structure of aceclofenac Table 4A.1 Interpretation...

Words: 996 - Pages: 4

Premium Essay

Analytical Chemistry Thesis

...pollutants emitted from industrial and domestic sources. Pesticides can often enter water through sewage plants, sewer overflows, surface water runoffs, and leaching from agricultural applications or even through improper disposal. Regulatory issues and public concerns over pesticide residues in water supplies have been increasing, and it is important to develop a reliable method for routine test of pesticides in water supplies. Presented here is a novel solid phase extraction technique using disposable pipette extraction (DPX) followed by chromatography and mass spectrometry analysis for determination of pesticides in water. The purpose of the current study was to evaluate the suitability of high pressure liquid chromatography with UV detector (HPLC/UV), gas chromatography flame ionization detector (GC/FID), gas chromatography electron capture detector (GC/ECD) and gas chromatography mass spectrometry (GC/MS) for pesticides analysis in water. It was found that GC/FID was not suitable due to its low sensitivity and poor selectivity. GC/ECD is the most sensitive method for pesticide analysis;...

Words: 7995 - Pages: 32

Premium Essay

Spectroscopy

...Week 7 SPECTROSCOPY Learning outcomes • Describe what a photon is, and calculate the energy, frequency and wavelength of a given photon in relation to electromagnetic radiation. • Describe the process where a UV/visible photon is absorbed by an atom or molecule. • Interpret the data from UV/visible spectrum: – Complete dilution calculations in order to produce a standard curve. – Determine an equation for the line of best fit to a linear set of data. – Use the equation, y = mx + c in order to determine the concentration of an unknown solution. Electromagnetic Radiation • Electromagnetic radiation consists of an oscillating electric and magnetic field that carries energy through space at the speed of light, c, Amplitude c=  ×  C = speed of light, 3.00 x108 m/s  = wavelength, m  = frequency, (number of waves per second) s-1 Maxwell ‘s description of behaviour of light Longer  Lower  Wavelength (SI unit : meter) Frequency (SI unit : second-1/ Hz) 1 MHz = 106 Hz 1 GHz = 109 Hz 1 THz = 1012 Hz 1 m = 10-6 m 1 nm = 10-9 m 1 pm = 10-12 m Example 7.1 The wavelength of the green light from a traffic signal is centered at 522 nm. What is the frequency of this radiation? Solution c=  ×  ������ =  Try this: 7.9 The average distance between Mars and Earth is about 1.3 x 108 miles. How long (in minutes) would it take TV pictures transmitted from the Viking space vehicle on Mars’ surface to reach earth? (1 mile = 1.61 km) ������������������������������������������������...

Words: 1408 - Pages: 6

Free Essay

Production of Bioethanol from Glycerine Using Saccharomyces Cerevisiae

...Application Note: 51853 Determination of Free Glycerol in Biodiesel with the Evolution Array UV-Visible Spectrophotometer Nicole Kreuziger Keppy, Gordon Bain, Ph.D., Michael W. Allen, Ph.D., Thermo Fisher Scientific, Madison, WI, USA www.analiticaweb.com.br/newsletter/08/AN51853_Biodiesel_UV.pdf Introduction Key Words • ASTM Standards • Biodiesel • EN Standards • Free Glycerol • Transesterification • UV-Visible Spectroscopy The consumption of biodiesel as an alternative fuel source is on the rise. Biodiesel is used in its pure form or added to fuel blends to power automobiles and heat homes. Biodiesel is typically produced from vegetable oils and animal fats by adding an alcohol, usually methanol, to triglycerides in a transesterification reaction as shown in Figure 1 below. The product of the transesterification reaction is the biodiesel fuel itself in the form of fatty acid methyl esters (FAME) and a glycerol by-product.1 Free glycerol affects the quality of biodiesel and its content is regulated by industry standards. In this experiment, the amount of free glycerol in biodiesel is measured with a Thermo Scientific Evolution Array UV-Visible spectrophotometer for conformance to American (ASTM)2 and European (EN)3 standards for free glycerol using the Quantification Mode of the Thermo Scientific VISIONcollect software. A solvent solution containing a 1:1 ratio of deionized water and 95% ethanol, and a reference solution of 0.036 mg/mL...

Words: 990 - Pages: 4

Free Essay

Calibration

...will be using a Zinc lamp on the flame atomic absorption spectrophotometer. The lab assistant will demonstrate the software and the procedures for preparing the instrument. Each group will run a series of calibration standards to generate a calibration curves. The concentrations are given below. Then, one of the "unknown" samples (A or B) should be run. The printer will generate a report of the concentration of Zinc in the sample. We have to be sure which sample we are used. Our laboratory report should describe the functions of the AAS and describe the procedures for preparing the instrument for use. The report should include a printout graph of the calibration curve that we have made. We also needed to indicate the data points for the calibration standards and our sample. INTRODUCTION Atomic absorption spectroscopy (AAS) determines the presence of metals in liquid samples. Metals include Fe, Cu, Al, Pb, Ca, Zn, Cd and many more. It also measures the concentrations of metals in the samples. Typical concentrations range in the low mg/L range. In their elemental form, metals will absorb ultraviolet light when they are excited by heat. Each metal has a characteristic wavelength that will be absorbed. The AAS instrument looks for a particular metal by focusing a beam of UV light at a specific wavelength through a flame and into a detector. The sample of interest is aspirated into the flame. If that metal is present in the sample, it will absorb some of the light, thus reducing...

Words: 2383 - Pages: 10

Free Essay

Identification of Statins in Rice Fermented with Monascus Species from the Museum of Natural History

...available statin drugs, this study idnetified the statins that can be produced by fermentation of rice by Monascus strains available at the Museum of Natural History (MNH). It also shall serve as a set-off point for further studies regarding the use of fermentation products for treatment of certain ailments, such as heart disease. A total of eight (8) isolates available at the Museum of Natural History (MNH) of the University of the Philippines – Los Baños (UPLB) were used for the production of angkak rice. Among these eight (8) strains, six (6) strains were able to produce the three commercially available statins used as standards, namely: simvastatin, rosuvastatin and atorvastatin. Ethanolic extracts of the angkak rice were measured through UV spectrophotometer and plotted against a standard curve. All the...

Words: 3052 - Pages: 13

Free Essay

Introduction for Biochemistry Lab -Spectrophotometry

...Introduction Every chemical compound absorbs, transmit or reflect light over a certain range of wavelength. In this experiment one of the main purpose is to introduce students to the quantitative and qualitative study of spectrophotometry. Spectrometry is the measurement of light that is absorbed or transmitted by the sample solution (Vo, n.d). It is one of the most commonly used methods in various fields to study the quantitative analysis of light that passes through a solution. Meanwhile, the use of spectrophotometer to determine the extent of absorption of wavelengths of visible of light is known as colorimetry. The intensity of light that passes through the sample as well as the concentration of the substance is measured by a spectrophotometer. The instrument consists of some basic components: Radiation light source (tungsten for visible region and deuterium lamp for ultraviolet region), wavelength selector (to isolate a desired wavelength form the source), a cuvette or container to hold the sample solution and a detector (photometer) which delivers the signal in a display device. The voltage output is determined by the difference in the amount of light absorbed by the sample. The fraction of light that passes through the solution is called transmittance, T and the proportion of light absorbed is the absorbance, Abs. Thus, the value of transmittance can be calculated by using the following formula: Transmittance, I = It/ I0 where, It = intensity of...

Words: 716 - Pages: 3